61
8. Operation
There are two step sizes:
Fine
and
Coarse
.
In the
Fine
mode, the value is adjusted to the
objective in use. The values are pre-defined.
The settings can be changed with the Software
Leica DMControl, module: Fine Tuning.
In the
Coarse
mode, the value is the same for all
objectives. Coarse represents the maximum
velocity.
Note:
When you assign a certain step size to an
objective, this does not only affect the Z-drive
but also the stage-step size that is assigned to
this objective when selecting the
Precise
mode.
Switch between Fine and Coarse via:
LeicaScreen - navigation item
Variable function keys at the stand and at
the SmartMove
Software Leica DMControl, module:
DMOperation
8.4 Tubes
Note:
Close any unused tube openings, as otherwise
stray light can interfere with observation.
Note:
Make sure that the connector cable is plugged
in on the motorized tube (67.1).
Adjusting the Viewing Distance
• Adjust the viewing distance of the eye-
pieces so that a congruent total image is seen
(Fig. 67).
Adjusting the Viewing Angle
• For the AET22 and EDT22 ergotubes, the view-
ing angle can be adjusted by tilting the bin-
ocular viewer in the range of 5° - 32° (Fig. 68)
1
↔
↔
↔
↔
↔
Fig. 67
Tube setting
↔
↔
↔
↔
↔
Personal eyebase settings
1
Motorized tube connection