Leica DM750 P User Manual
Safety Regulations 15
Electrical data and ambient conditions
Environment
Temperature for use
+10 °C … +40 °C
Storage temperature
-20 °C … +52 °C
+50 °F … +104 °F
Manipulation shock
25 mm on 50 mm hard wood
Transport shock (unpacked)
100 g / 6 ms
Transport shock (packed)
800 mm free fall
Transport vibrations (unpacked)
5–200 Hz / 1.5 g
Atmospheric pressure "in use" and for storage 500–1,060 mBar
Humidity "in use" and for storage
20–90 %
Installation Category II (Overvoltage Category)
Pollution degree 2
Fuse Replacement
Unplug the instrument before changing any fuses. The
Leica DM750 P has two fuses, which are located behind the
power cord receptacle.
•
Use only fuse type: 5×20 mm, 1 Amp/250 V, Quick acting
(# 13RFAG30003)
Electrical data
Input: 100–240 V, 50/60 Hz, 5 W (3 W LED)