Electrical data and ambient conditions
In order to maintain this condition and to ensure safe operation,
the user must follow the instructions and warnings contained in
this instruction manual.
Environment
Temperature for use
+10 °C … +40 °C
Storage temperature
-20 °C … +52 °C
+50 °F to +104 °F
Manipulation shock
25 mm on 50 mm hard wood
Transport shock (unpacked)
100 g / 6 ms
Transport shock (packed)
800 mm free fall
Transport vibrations (unpacked)
5–200 Hz / 1.5 g
Air pressure during use and storage:
500–1060 mbar
Humidity during use and storage:
20–90%
Installation Category II (Overvoltage Category)
Pollution degree 2
Fuse Replacement
Unplug the instrument before changing any fuses. The Leica
DM750 has two fuses, which are located behind the power cord
receptacle.
Only use the following fuse types: 5×20 mm, 1 A/250V, fast-acting
fuse (# 13RFAG30003)
Electrical data
Input: 100-240 V, 50/60 Hz, 5 W (3 W LED)
General safety notes
This instrument of safety class 1 has been built and tested in accordance
with the following safety requirements for electrical equipment for measu-
rement, control, and laboratory use:
EN 61010-1: 2002-08
EN 61010-2-101: 2008-06
IEC 61010-1: 2010-06
EN 61326-1: 2006-10
EN 61326-2-6: 2006-10
EN 55011: 2007+A2: 2010-05
EN 60825-1: 2008-05
IEC 60825-1: 2007-03
LED Class 1
Summary of Contents for DM750
Page 1: ...Leica DM750 Manual...
Page 5: ...Leica DM750 Manual 5 Safety regulations...
Page 17: ...Leica DM750 Manual 17 The Leica DM750...
Page 20: ...Leica DM750 Manual 20 Ready...
Page 30: ...Leica DM750 Manual 30 Set...
Page 43: ...Leica DM750 Manual 43 Go...
Page 45: ...Leica DM750 Manual 45 Care of the Microscope...
Page 48: ...Leica DM750 Manual 48 Troubleshooting...
Page 51: ...Leica DM750 Manual 51 Dimensions...
Page 52: ...Leica DM750 Manual 52...
Page 53: ...Leica DM750 Manual 53...
Page 54: ...Leica DM750 Manual 54...