If your Leica DM750 is fitted with a field diaphragm for Koehler Illu-
mination, use the following procedure to assure condenser cente-
ring and focus.
1. Connect the Koehler field diaphragm to the base of the microscope
so that the diaphragm leaves are in the field of view when you look
through the eyepieces.
2. Focus the leaves of the field diaphragm using the condenser focusing
knob on the left side of the stage mount.
Koehler Configuration
Close the Koehler field diaphragm
Condenser focusing knob
Close the Koehler field diaphragm
Leaves are in sharp focus