Leica DM750
Manual
32
If you have purchased a standard Leica
DM750 configuration, the condenser has
already been pre-centered by Leica Microsys-
tems.
1. Open the condenser aperture by rotating
the knurled ring on the condenser to the
right.
Make sure that the condenser is in the
highest position.
2. Write an "X" on a piece of paper in the size
of a business card and place it onto the
light output of the microscope stand in
such a way that the "X" is centered over the
illumination.
3. Look at the X through the eyepieces and
center it in the field of view by tightening
the setscrews with the tool included in the
delivery package (or, for a Leica DM750
with Koehler illumination, tighten the
knurled screws).
Complete Condenser Centering
Not centered
Centered