2. Safety NoteS
2. Safety Notes
2.1 General
safety notes
Caution!
Make sure to follow the safety notes in the Leica DMi8
User Manual.
Caution!
Also see the instructions and warnings contained in this
Installation Manual.
Caution!
The instruments and accessories described in this In-
stallation Manual have been safety-tested and checked
for possible hazards.
The responsible Leica representative or the main plant
in Wetzlar must be consulted whenever the device is
altered, modified or used in conjunction with non-Leica
components that are outside of the scope of this manual!
Unauthorized alterations to the instrument or noncom-
pliant use shall void all rights to any warranty claims!
2.2 Technical data
Compact Leica CTR electronics box
For indoor use only.
Supply voltage:
100 – 240 VAC
Frequency:
50 / 60 Hz
Power consumption:
max. 150 VA
Fuses:
3.15 A, slow-blowing,
breaking capacity H, 250 VAC
Size: 5x20 mm
Ambient temperature:
15° - 35°C
Relative humidity:
90% up to 30°C,
non-condensing
Protection class:
I
Overvoltage category:
II
Pollution degree:
2
Advanced/A Leica CTR electronics box
For indoor use only.
Supply voltage:
100 – 240 VAC
Frequency:
50 / 60 Hz
Power consumption:
max. 290 VA
Fuses:
6.3 A, slow-blowing,
breaking capacity H, 250 VAC
Size: 5x20 mm
Ambient temperature:
15° - 35°C
Relative humidity:
90% up to 30°C,
non-condensing
Protection class:
I
Overvoltage category:
II
Pollution degree:
2
6