29
7.2 Adjusting the scoring pressure
Raise the scoring head clompletely as shown in
Fig.36
and place it on the table
upside down.
By turning the screw
(Fig.39)
with the 2 mm Allen Key (provided)
clockwise
the pressure
can be
increased
,
counter clockwise
the pressure can be
decreased
.
After turning the screw in the required direction (a quater to a half turn)
replace the scoring head and make a test score on a piece of glass.
Repeat until the pressure is correct.
The pressure is correct
(Fig.40)
- one faint smooth line without glass splinters.
The pressure is too strong
(Fig.41)
- a groove and glass splinters can be seen.
The pressure is not enough when the scoring line is faint and interrupted.