Troubleshooting
4
Troubleshooting
Cyclone
Problem
Possible Cause(s)
Suggested Remedies
A scan runs without visible
laser radiation.
Laser not activated.
Check that there are no
objects within the minimum
monitoring range. Minimum
monitoring range, for initial
"Laser Safety" scan is 30 cm.
Cyclone reports that the
temperature is too low.
Temperature is too
low.
Start scanning, e.g. perform
some preview scans, as the
activity helps warm the
HDS6000.
Cyclone reports that the
temperature is too high.
Temperature is too
high.
Leave the HDS6000
powered. Wait until the
active cooling in the scanner
stops before resuming scan-
ning.
Problem
Possible Cause(s)
Suggested Remedies
Cyclone can not connect
with HDS6000.
Check the connection. Check the data cable connec-
tion.