Functional Check
57
Functionality check
Before any tests can be carried out it is vital to check the status of the unit, its batteries
and basic functionality.
The following list is used to achieve this.
1. Inspection
Once the general condition of the DIGICAT is established the self test can then be used to
indicate the basic unit function and state of the batteries.
2. Audio / Visual display test
Upon depressing the trigger the DIGICAT should test the display and speakers by illuminating
each segment in the bar display followed by the mode LEDs and Depth display.
All LEDs must be operative and an audible output must be heard.
3. Battery / Functional self check
If there is no response when the trigger is activated or the low battery illuminates (or
flashes) after the Audio / Visual Display test, the Batteries will have to be replaced. Use alka-
line batteries, replace all of the batteries at the same time.
•
Casing
The Casing should be free of significant damage.
•
Labels
Body labels must be legible and intact. Display label must be
free of damage and tears.
•
Battery Hatch
The Hatch must lock into place.
•
Battery Holder
All the battery contacts and springs on the Holder must be free
of corrosion and the holder in good condition.
•
Battery Contacts
The Battery pack contacts must be free of corrosion.
Summary of Contents for LASER ALIGNMENT DIGICAT 100
Page 1: ...Leica DIGICAT 100 200 English Français Español ...
Page 2: ......
Page 3: ...Leica DIGICAT 100 200 User Manual Version 3 1 English ...
Page 53: ...Safety Directions DIGICAT 100 200 51 Labelling Cable Avoidance Tool ...
Page 69: ...Index DIGICAT 100 200 67 ...
Page 71: ...Leica DIGICAT 100 200 Manuel de l utilisateur Version 3 1 Français ...
Page 122: ...52 DIGICAT 100 200 Consignes de sécurité Etiquetage Cable Avoidance Tool ...
Page 139: ...Index DIGICAT 100 200 69 ...
Page 141: ...Leica DIGICAT 100 200 Manual de empleo Versión 3 1 Español ...
Page 191: ...Instrucciones de seguridad DIGICAT 100 200 51 Rótulo Cable Avoidance Tool ...
Page 209: ......