CONTINOUS SMART TARGETING FUNCTION (Automatic alignment)
The continuous Smart Targeting function is selected, but does not start?
Pressing Sensitivity key and Smart Targeting key simultaneously is sometimes difficult to handle. If
the Sensitivity key is pressed first, it changes only the sensitivity and you see that the sensitivity
icon changes in the display and the continuous function is not activated. To overcome this effect,
we recommend to press first the Smart targeting key and shortly afterwards additionally the sensi-
tivity key. If you release the keys, 2 small blinking arrows appear on the display near the Smart
Targeting icon.
ADDITIONAL INFORMATION
MANUAL ROTATION
The device is rotated manually with the knob and it does not move right from the begin-
ning?
It is a normal behaviour, that it takes few manual turns to move finally the laser beam. The ML de-
vice is focused on targeting laser beams at large distances and this
“slow down effect” prevents
displacing the beam due to small accidental pushes.
BASE PLATE IS MOVING
The base plate is not fixed properly - is this a mistake?
The base plate can be moved manually. This is on purpose as this is the movement for the +/-4°
rotation.
LEVELLING PROCEDURE
The vertical lines switch off, if the device is levelling?
The high accurate electronic levelling needs power and therefore the vertical lines can shut off dur-
ing the levelling procedure. This is to save power and to speed up the levelling. After finalizing, the
lines shut on automatically and you can work as usual.
+
=