CONTROLLED COPY 

 Centre for Biological Engineering 

 

Standard Operating Procedure    

 

 

 

 

 

 

SOP194 

 

Title:  Use and maintenance of Leica RM2125 RTS rotary microtome.

 

 

Location: 

H27 

 

 

Version 001 

Effective Date: 24.05.2021 

Review 24.05.2023 

 

Written by: Sotiria Toumpaniari 

 

Reviewed by:  

Approved by: C.Kavanagh 

 

 

Page 8 of 11 

 

CAUTION: The handwheel continues turning if it has been rotated very quickly and then released – this can 

cause crushing or other injuries! 

8.  Set the desired section thickness using the section thickness setting knob (26) at the front of the 

microtome on the right, or check the set value in the viewing window (25). The red indicator (24) 

indicates the selected section thickness (on the scale).  

9.  Use a different area of the cutting edge for trimming than for sectioning. 

10. To do so, move the knife holder laterally on the knife holder base accordingly (see Chap. 5.9.7) or, 

when using the knife holder base without lateral movement, move the knife or disposable blade in 

the knife holder.  

11. For sectioning, turn the handwheel (15) evenly in a clockwise direction.  

12. Pick up the sections and mount them on microscope slides. 

 

5.3.8.  Changing specimens  

Note: 

Lock the handwheel and cover the knife edge with the knife guard prior to any manipulation of knife or 

specimen, as well as prior to changing specimens and during all work breaks!  

1.  Move the specimen to the upper end position by turning the handwheel and engage the handwheel 

lock.  

2.  Cover the sectioning edge with the knife guard.  

3.  Remove the specimen from the specimen clamp and mount a new specimen.  

4.  Run the object clamp with the coarse feed back far enough until the new specimen can start being 

cut. 

5.4. Cleaning and Maintenance 

5.4.1.  Cleaning 

 

Note: 

Always remove the knife / blade before detaching the knife holder from the instrument. Always put the 

knives back into the knife case when not in use! Never place a knife anywhere with the cutting edge facing 

upwards and never try to catch a falling knife! When using cleaning agents, observe the manufacturer's 

safety instructions and the laboratory regulations valid in the country of use. Do not use any of the following 

for cleaning the outside surfaces of the instrument: alcohol, detergents containing alcohol (window cleaner!), 

abrasive cleaning powders, solvents containing acetone or xylene. Xylene or acetone will damage the 

finished surfaces! Ensure that liquids do not enter the interior of the instrument during cleaning!Move the 

specimen to the upper end position by turning the handwheel and engage the handwheel lock.  

Before each cleaning, carry out the following preparatory steps:

  

  Move the specimen clamp to the upper end position and activate the handwheel lock. 

  Remove the blade from the knife holder and insert it in the receptacle at the bottom of the 

dispenser, or remove the knife from the knife holder and put it back in the knife case. 

  Remove knife holder base and knife holder for cleaning. 

  Remove the specimen from the specimen clamp. 

  Remove section waste with a dry brush. 

  Remove specimen clamp and clean separately. 

Instrument and outside surfaces: 

  If necessary, the varnished outside surfaces can be cleaned with a mild commercial household 

cleaner or soap water and then be dried with a moist cloth. 

 

Summary of Contents for RM2125RTS

Page 1: ...laboratory staff shall i Work in accordance with the instructions for operation and maintenance detailed in this SOP Risk Assessment and the operator manual ii Carry out the routine inspection cleani...

Page 2: ...the system The Leica RM2125 RTS is a manually operated rotary microtome for creating thin sections of specimens of varying hardness for use in routine and research laboratories in the fields of biolo...

Page 3: ...lly or vertically as required 6 To clamp the cassette release the lever Note Before clamping the cassette into the universal cassette clamp remove excess wax on the outside of the cassette to ensure t...

Page 4: ...the Allen screw to release the clamp by turning Allen key No 4 9 counterclockwise 4 When using the knife holder base with lateral movement rotate the lever on the right side of the knife holder base c...

Page 5: ...ies to the rocking mode during which the specimen is trimmed by rocking motion no complete rotation of the handwheel Carry out rocking mode ONLY in the sectioning phase NEVER in the retraction phase 5...

Page 6: ...clockwise see curved arrow means the specimen advances Turning the coarse feed wheel clockwise means the specimen retracts away from the knife 2 Red dot at 12 o clock turning the coarse feed wheel clo...

Page 7: ...has 3 click stops 0 m 10 m and 50 m The points 23 mark the two trimming stages 10 m 50 m 1 To activate the trimming function press the lever downwards into the desired position and keep it pressed do...

Page 8: ...cimen clamp and mount a new specimen 4 Run the object clamp with the coarse feed back far enough until the new specimen can start being cut 5 4 Cleaning and Maintenance 5 4 1 Cleaning Note Always remo...

Page 9: ...ing the parts from the drying chamber 65 C Wearing safety gloves is recommended 8 After cleaning the moving parts apply a thin coat of drive part oil to them 9 Reassembly takes place in reverse order...

Page 10: ...e holder N The clamping lever 89 on knife holder E Shaft 90 of the clamping lever of the cassette clamp see Fig 36 Special Safety Note Equipment contains sharp edges Ensure you are wearing PPE have be...

Page 11: ...pproved by C Kavanagh Page 11 of 11 6 DOCUMENTATION 6 1 Equipment Maintenance Log 6 2 Microtome Training Agreement This form can be found on the Loughborough University LEARN module for the Centre of...

Reviews: