17
Calibration
Rugby 200-3.2.1en
Calibration
)
The Rugby is calibrated to the
defined accuracy specification at the
factory. It is recommended to check
your laser for calibration upon receipt
and periodically before using the laser
to ensure calibration is maintained. If
your laser requires calibration, send it
to your nearest authorized service
center, or calibrate the laser using the
following procedure.
)
Do not enter this mode or
attempt calibration unless you plan to
change the calibration. Calibration
should be performed only by a
qualified individual that understands
basic calibration principles.
Checking Level Calibration
1 To check level calibration of your
Rugby Laser, place the unit on a
flat, level surface or tripod approxi-
mately 100 ft (30 m) from a wall.
2 Align Axis one of the Rugby so that
it is square with the wall. Allow the
unit to self-level completely
approximately one minute after the
unit begins to rotate), and then
mark the position of the beam
(position 1).
3 Rotate the Rugby 180°, allow it to
self-level and mark the opposite
side of the first axis (position 2).
4 Align the second axis of the Rugby
by rotating it 90° so that this axis is
now square with the wall. Allow the
unit to self-level completely, and
then mark the position of the beam
(position 3).
5 Rotate the Rugby 180°. allow it to
self-level and mark the opposite
side of the second axis (position 4).
The Rugby is within its calibration
specification if the four marks are
within ±1/16” (1.5 mm) from center.
Rugby200-16 Axis 1_1.eps
Axis 1 (Position 1)
Rugby200-17 Axis 1_2.eps
Axis 1 (Position 2)
To Enter Calibration Mode
1 With the power off, press the Power
Button once to turn on power.
Then, press the head speed button
to set the head speed at its fastest
setting (10 rps).
2 Press and hold the Up and Down
Arrow Buttons.
While holding these
buttons, press and release
the power button. Then,
release the arrow buttons.