20
Troubleshooting
Rugby 200-3.2.1en
Troubleshooting
Symptom
Possible Causes and Solutions
The Rugby 200 does not
turn on.
The battery charge is low or dead.
•
If the Rugby 200 has alkaline batteries, remove
the battery pack from the base of the Rugby
and replace all four batteries with fresh D-cell
alkaline batteries.
•
If the Rugby 200 has a rechargeable) NiMH-
battery pack, plug the battery charger into the
Rugby's charge jack and then connect it to a
standard electrical outlet. Always plug the
battery charger into the Rugby's charge jack
first and then plug it into the electrical outlet,
otherwise the Rugby's charge jack might emit
sparks.
The Rugby 200 is working,
but it does not self-level.
The Rugby 200 must be in automatic mode to
self-level.
•
In automatic the emission indicator will blink
until the laser has completed self-leveling.
•
In manual mode the Level Warning Indicator
will flash at 5 Hz, and the Rugby will not self-
level.
The Rugby 200 has
stopped working and the
H.I. LED indicator is
flashing.
The Rugby 200 is displaying an elevation alert.
•
The H.I. Indicator will turn on solid when the H.I.
is activated and will flash when the laser has
moved and an alert condition exists.
•
Press the H.I. or Auto/Manual Button to stop the
alert. Check and adjust the elevation of the
laser, then press the H.I. button to reactivate
the H.I. Alert function.
The Laser Emission, Low
Battery and Manual Mode
Indicators are flashing
sequentially. (Out of Level
Indication)
The Rugby 200 has reached a servo limit or is
outside of its self-leveling range.
•
In automatic mode the Rugby 200 must be set
up within ± 5° of its upright position to be within
the self-leveling range.
•
Reposition the Rugby. If the LED’s continue to
flash, turn off the Rugby and try again.
•
In manual mode, reverse directions and
continue working.