Fig. 15
Spectroscopic detection of a system with 5 PMTs
Fig. 16
Slit and PMT settings control window
4.4.4. Setting the slits for the spectral separation of the emitted light
The sliders in the beam path window underneath the spectral view tool in Fig. 16 (S1 to 3)
represent the three slits of the spectroscopic detection unit. By left-clicking and moving
them, they can be shifted to shorter or longer wavelengths without changing the slit width.
By clicking and moving the left or right edge of the bar, the appropriate width of the slit can
be changed. The sliders can be used to adjust the position and width of each slit
accurately down to the nm. Right-clicking the sliders will open a little window that indicates
the settings of each slit in nm (see Fig. 16). This can also be used to set the spectral
separation.