background image

_____________________________________________________________________________________  

 

Rolly Wiegand – CALM Leica SP5 manual

 

 

22

 

dimensions of the entire image. The approximate XY-pixel dimensions at maximum 
resolution using a sampling rate that fulfils the Nyquist criterion is 45 nm and further 
zooming results in over-sampling. 
 
 

   Fig. 29 

‘Zoom’ navigator 
 
This function will zoom in on the centre of the entire field. However, if you want to image 
part of the entire field that is not in the centre, you can use the arrow buttons (see Fig. 29, 
circle) to shift the area you have zoomed in to the desired position within the entire field. 
 
4.7.5. Averaging 
This function allows scanning the same optical section several times, as can be defined by 
the user, and automatically calculates the average image of several scans, which 
significantly reduces the background noise, and thus improves the signal-to-noise ratio. 
For instance, if a pixel from a ‘real’ signal in the image records the intensities of 120, 125, 
115 and 120 in four consecutive scans, the average would be 120 and would constitute a 
bright feature in the image, whereas a single random photon hit in a different pixel in only 
one of the four scans of the same image with an intensity of 40 would have an average of 
10 and thus a low intensity feature. However, more averaging obviously causes more 
photobleaching, in particular if large Z-stacks are acquired. It is therefore recommended 
not to increase averaging to more than four scans unless the fluorescent label is very 
stable. If photobleaching still is a problem, averaging should be reduced or avoided at all. 
Bear in mind that the remaining noise in the image can be removed by post-acquisition 
image deconvolution and appropriate hardware and software is available in the CALM 
facility. Line averaging is the faster option, but for fixed samples, averages between 
frames might provide a very short recovery time from photobleaching. 
 
4.7.6. Accumulation 
Similar to the averaging function, this mode allows accumulating the signal from several 
scans into one image without averaging and can be used if the photon count from weak 
fluorophores is very low. However, bear in mind that unlike in the averaging mode, in the 
accumulation mode the noise will be added up as well and will multiply by the number of 
scans.  
 
4.7.7. Scan field 
By tilting the scanning mirrors of the system, the scan field can be rotated by +/- 180

˚

. This 

is a useful tool if the object to be imaged can only be fully fitted into the scan field by 
rotating it. This function will slightly increase the overall scanning time. 
 
4.8. Acquisition of Z-stacks  
For post-acquisition image deconvolution and accurate quantitation, it is always 
recommended to image entire cells or tissue sections in 3 dimensions rather than 
representative single sections. Programming the acquisition of a Z-stack requires to define 
the start and end section to be imaged as well as the Z-step (or Z-interval), which defines 
the stepping of the Z-galvo between the different optical sections.  

Summary of Contents for SP5C

Page 1: ...tute College for Medicine and Veterinary Medicine University of Edinburgh Manual for the Leica SP5C Spectral Confocal Laser Scanning Microscope Version 1 11 September 2008 Written and illustrated by R...

Page 2: ...signal 12 4 4 4 Setting the slits for the spectral separation of the emitted light 13 4 4 5 Adjusting the PMT settings 14 4 4 6 Additional transmitted light channel 16 4 4 7 Saving and loading of beam...

Page 3: ...Emission Fig 1 1 3 Wait until the PC has finished booting up 1 4 Log into Windows XP using the general account TCS user which is not password protected 1 5 IMPORTANT Make sure that the objective turre...

Page 4: ...Selection of the objective lens The LAS AF software will open showing the Acquisition page Select the objective you want to use in the Beam Path window see Fig 3 IMPORTANT Never change the objective b...

Page 5: ...d mark indicates the corrective settings for 37 C Fig 4 Choose the lens you want to use by matching the refractive index RI of the immersion medium with the medium embedding your sample oil lens with...

Page 6: ...erslip you should be using should be of thickness 1 130 150 m mean thickness The configuration with the motorised stage is not particularly suitable for live cell imaging and we recommend the Zeiss LS...

Page 7: ...ragm the illuminated field will decrease to the area chosen The fluorescence light path is also fitted with a shutter that allows to block illumination of the sample and thus to protect it see the yel...

Page 8: ...ee Fig 10 and allows you to conveniently find your initial marked focal plane again IMPORTANT Never push the objective lens against the coverslip because this might damage the front lens and might req...

Page 9: ...s switched on The display at the front of the power supply see Fig 5 shows the total usage time of the present bulb If this reaches 220 hours please inform the CALM facility staff by e mail so that th...

Page 10: ...on the display at the front see Fig 10 Fig 10 Display on the front of the microscope stand You can find more details about the DMI6000 microscope stand in the Leica handbook and a copy of the pdf file...

Page 11: ...Laser Excitation wavelength nm 405 Blue Diode 405 Argon 458 476 488 496 514 Helium Neon 1 543 Helium Neon 2 594 Helium Neon 3 633 4 2 Setting the USB Control Panel In the configuration tab open the w...

Page 12: ...r you can define your own settings and save them for later re use Before you start choosing your settings for a multi channel image acquisition you have to decide whether to scan the channels simultan...

Page 13: ...equential combined channel 1 and 3 in the first and channel 2 in the second sequential scan This combined method involves 2 sequential scans During the first scan channels 1 and 3 are recorded Cross t...

Page 14: ...2 Spectrum view tool In order to assist and facilitate the definition of the beam path settings the software provides a visual tool that shows the excitation lines as well as the emission spectra of a...

Page 15: ...Fig 16 S1 to 3 represent the three slits of the spectroscopic detection unit By left clicking and moving them they can be shifted to shorter or longer wavelengths without changing the slit width By cl...

Page 16: ...The accurate adjustment of each PMT with regard to each fluorophore signal is absolutely crucial for a valid image acquisition For instance gain settings that lead to saturation and thus clipping of...

Page 17: ...4 5 2 check the off set settings again and correct if necessary 4 4 5 2 Setting the PMT gain Whilst continuing to scan adjust the voltage gain setting on the control console the range is 0 1250 V and...

Page 18: ...f specificity and also contain out of focus light they often give valuable spatial information such as cell outlines or the position of the cell nucleus To access the settings click on Additional Chan...

Page 19: ...needs to be re adjusted and if you are not familiar with the condenser unit please inform any CALM staff member Fig 22 Both channels with correct settings 4 4 7 Saving and loading of Beam Path setting...

Page 20: ...d 3 spectrally different channels at the same time By activating the desired PMTs see 4 4 5 and adjusting all necessary settings for the beam paths the simultaneous scanning can be set up easily 4 5 2...

Page 21: ...ber of sequential scans To save time you can scan more than one spectral channel in one of the sequential scans as long as the excitation and emission spectra do not overlap see the third option of th...

Page 22: ...most possible combination of x y z t time and wavelength sweep are available for image acquisition in 2 3 or 4 dimensions 4 7 Setting the scanning parameters The window panel below the Acquisition mo...

Page 23: ...to be reduced However longer dwell times i e slower scan speed also increases the bleaching and this might become a problem with dim samples that are at the same time prone to quick bleaching If as i...

Page 24: ...herefore recommended not to increase averaging to more than four scans unless the fluorescent label is very stable If photobleaching still is a problem averaging should be reduced or avoided at all Be...

Page 25: ...on in the sample which increases with the thickness of the sample that needs to be penetrated Because the entire specimen illuminated by the light cone from the objective lens will suffer from photobl...

Page 26: ...ing effects of the whole specimen caused by the repeated scanning should be taken into account If bleaching is a problem avoid averaging and rather rely on removing the increased noise in the image by...

Page 27: ...y clicking the Live button Then use the dials for the Smart Gain and the Smart Off set to optimise your settings see Fig 19 Click the Live button which now shows as Stop again to stop scanning Check a...

Page 28: ...o your destination folder on the hard drive in between different scans We strongly recommend that you do this after each scan also renaming each new file accordingly 5 1 Saving images and experiments...

Page 29: ...er by annotating each file To achieve this right click the appropriate file name and choose File properties This opens a window showing all metadata of the file which are accessible at any time post a...

Page 30: ...are packages for image deconvolution and analysis are available in the bioinformatics room C2 38 You find more information about this on the CALM website www calm ed ac uk 7 Shut down procedure Please...

Page 31: ...can occur and recommendations of how they can be fixed F In any case any serious problem should be logged reported to the CALM facility staff and an error report should be generated within the softwar...

Page 32: ...Fig 11 Check whether the required lasers have been selected in the Beam Path Settings Check whether the arm that carries the tungsten light source for transmitted light is in its correct position i e...

Reviews: