background image

 

To view the specimen in bright field (BF), 4 things to do:  turn the power wheel # 11 on to 
about 4V 

(for white light at 3200 K around 8V, to be used with NDF) 

and the lollypop 

shutter above the field diaphragm in down position

 Knob #12 operating a mirror in the 

condenser light path should be on VIS = visual and MicCtrl on Visual.  

 

Focus on the specimen with the knob “Focus” ; step S3 is coarse, step S2 is medium, Step 
S1 is fine. 

 

Apply Koehler illumination by closing diaphragm #13a and adjusting the level of the 
condenser at knob # 13b to obtain a sharp projection of the edges of this diaphragm in the 
oculars.  The illumination path can be centered with the two rings #13c. 

 

Set #15 on BF. If more suitable for a certain objective 63x oil or 40/63 x).  

 

Knob #22 should be pushed inward 

 

Adjust diaphragm #14 (from 0 to over 8 and Ph) to have just enough contrast. 

 

Light can be dimmed with Neutral density filter (NDF) # 16. For bright field illumination 
pull polarization filter # 17 out of the light path. 

 

Turn the ribbed disc # 18 to position BF.  

 

Pull the (ICT P) analyzer (polarized filter) #19 out for bright field vision 

 

Keep #22 pressed in (half pulled out is 50% ocular-50% upper camera port vision; fully 
pulled out is 100% upper camera port vision). 

 

Ring #23a (display #23b) can be set in position 1x (open connection for bright field); UV 
for ultraviolet scan stand, B for phase contrast adjustment with Bertrand lens (not relevant 
in this microscope) , and scan for non-UV scanning 

 

Optional but easy is the function to set the z-focus level of the objective for later recovery. 
Press #27 ↑ for 3 s, focus and press #27 ↑ again. The lowest position of the objectives can 
be set in a similar way using the #25↓ switch (standard is around –150 micrometer). 

 

Differential Interference contrast settings

: Press the (ICT P) analyzer polarized filter 

#19 in for eye vision and out for laser scanning for this S23 condenser, turn either of the 
following prism on the #15 turret to the front: K10 for 63x gly and oil objectives, K8 for 
20 x and D1 for 10x. Switch polarizer/analyzer #16 into the light path. 

 

To watch wide-field fluorescence, choose the desired filter block with the arrows on 20d, 
and open shutter 20c (close the transmission light lolly pop) 

 

To go to the confocal imaging, choose scan from the MicCtrl icon, block the light of the 
transmission lamp, close the fluorescence lamp shutter. 

Summary of Contents for TCS SP2

Page 1: ...Brief manual how to start and close the Leica sp2 Confocal TCS SP2 AOBS system mounted on a DM IRE2 ...

Page 2: ...the scan head and lasers switches on Press red knob 3 on the cooling is functioning note that outside working hours an additional cooling should be switched on from the box on the corner just outside the room ask Liesbeth Switch the conventional 50 W HBO fluorescence lamp on by pressing once on 20a of EBQ 100 NEVER re ignate the lamp soon after having switched it off In general avoid any unnecessa...

Page 3: ...answering Yes to the question Initialize Stage Remember to perform xy stage initialization only with a 10x objective upright and to press the condenser arm backward if you use the xz galvo sample holder so that does not touch the condenser during the calibration move Click No to the question Initialize Stage if automated xy positioning is not applicable Choose the appropriate objective and immersi...

Page 4: ...out for bright field vision Keep 22 pressed in half pulled out is 50 ocular 50 upper camera port vision fully pulled out is 100 upper camera port vision Ring 23a display 23b can be set in position 1x open connection for bright field UV for ultraviolet scan stand B for phase contrast adjustment with Bertrand lens not relevant in this microscope and scan for non UV scanning Optional but easy is the ...

Page 5: ... laser choose passed through Neutral density filter means no neutral density filter Set PMTs for the desired wavelength ranges by mouse dragging trans active optional reflection is a special case Functions of the black knobs can be adjusted by right clicking on the lower bar menu choose configuration for menu choose sensitivity to adjust the coarsness of the knob e g Smart offset smart gain Use th...

Page 6: ...lly with the smart function knobs on the table panel or automatically with A Gain There is an extensive Help menu within the Leica Confocal software as well as a print of most help functions and tutorials in the confocal microscope room Acquired images are stored temporarily on a temp directory can be retrieved if the program crashes but not if the computer crashes therefore it is advisable to reg...

Page 7: ...since we had to remove F secure to save space on the 10G C drive USERS SHOULD REMOVE THEIR OWN DATA FROM THE HARD DISK WITHIN ONE WEEK to prevent congestion of the system Shut down Windows on the computer When ready switch 3 out Switch 2 out Switch 1 out D Good working practice Always reserve time through bookings science ru nl Do not hesitate to contact Liesbeth for assistance instructions if you...

Reviews: