145
FLIM Data Acquisition
If you would like to use other wavelengths for excitation, the filter wheel should be in the
Empty
position (
).
16.2.4
Changing the Pulse Frequency for Pulsed Diode Lasers (405, 440, 470,
640 nm)
The frequency is controlled using LAS AF. The corresponding signal is transferred from the
scan head via the pulse interface to the laser driver PDL or Sepia.
1. Make sure that the laser driver is set to external trigger.
2. In
Beam Path Settings
, in the area for controlling the pulsed lasers, select the desired
frequency in the pull-down menu (
). The default setting is 40 MHz.
Figure 129:
Beam Path Settings in LAS AF: Activating a pulsed diode laser and selecting the laser
pulse rate.
After changing the frequency, it may be necessary to change the time resolution of the data
acquisition in SymPhoTime. You can change the resolution in the oscilloscope window of
the SymPhoTime software as follows:
3. Select
Show Oscilloscope...
in SymPhoTime.
Summary of Contents for TCS SP8 SMD
Page 1: ...10 Living up to Life User Manual Leica TCS SP8 SMD for FCS FLIM and FLCS ...
Page 4: ...4 Copyright ...
Page 14: ...14 Contents ...
Page 18: ...18 Intended Use ...
Page 20: ...20 Liability and Warranty ...
Page 28: ...28 General Safety Notes ...
Page 32: ...32 Additional Notes on Handling the System ...
Page 44: ...44 System Overview and Properties ...
Page 60: ...60 SMD Components Figure 31 DSN 102 Dual SPAD Power Supply ...
Page 80: ...80 Safety Features ...
Page 102: ...102 Switching On the System ...
Page 116: ...116 LAS AF ...
Page 214: ...214 Changing the Specimen ...
Page 216: ...216 Changing the Objective ...
Page 238: ...238 Switching Off the System ...
Page 242: ...242 Repairs and Service Work ...
Page 244: ...244 Maintenance ...
Page 246: ...246 Disassembly and Transport ...
Page 248: ...248 Disposal ...
Page 254: ...254 Contact ...
Page 256: ...256 Recommended Literature ...
Page 266: ...266 Appendix Figure 225 Declaration of conformity ...
Page 268: ...268 Appendix ...
Page 269: ......