169
Summarized manual for FLIM experiments
1. Run a test measurement by selecting
Run FLIM Test
in LAS AF.
2. The
Measurement Preview
window opens in SymPhoTime (
window by selecting the
Stop
button.
3. You can now test the fading behavior of your specimen using the oscilloscope window
in the SymPhoTime software. Select
Show Oscilloscope...
4. Select the TCSPC tab
Histogram
, select
Start
and monitor the fading of the fluorescence
as shown in
Figure 149:
Fading of the fluorescence at too high of a laser repetition rate (A, 80 MHz), optimum laser
repetition rate (B, 40 MHz) and unnecessarily low laser repetition rate (C, 20 MHz).
5. In this window, you can select the time resolution per bin of the TCSPC histogram with
the slider under
Acquisition
.
6. Please confirm your selection by selecting
Apply
.
Repetition rate with pulsed diode lasers
In the supplied configuration, 40 MHz is the maximum repetition rate for
the pulsed diode lasers. 80 MHz is available only if the trigger on the laser
driver (PDL, SEPIA) is set to
internal
. This bypasses the laser settings in
LAS AF. In addition, all functions related to beam blanking (such as ROI
scanning or line-by-line sequential scanning) are deactivated. A change
of synchronization cable length will be necessary as well.
p
Summary of Contents for TCS SP8 SMD
Page 1: ...10 Living up to Life User Manual Leica TCS SP8 SMD for FCS FLIM and FLCS ...
Page 4: ...4 Copyright ...
Page 14: ...14 Contents ...
Page 18: ...18 Intended Use ...
Page 20: ...20 Liability and Warranty ...
Page 28: ...28 General Safety Notes ...
Page 32: ...32 Additional Notes on Handling the System ...
Page 44: ...44 System Overview and Properties ...
Page 60: ...60 SMD Components Figure 31 DSN 102 Dual SPAD Power Supply ...
Page 80: ...80 Safety Features ...
Page 102: ...102 Switching On the System ...
Page 116: ...116 LAS AF ...
Page 214: ...214 Changing the Specimen ...
Page 216: ...216 Changing the Objective ...
Page 238: ...238 Switching Off the System ...
Page 242: ...242 Repairs and Service Work ...
Page 244: ...244 Maintenance ...
Page 246: ...246 Disassembly and Transport ...
Page 248: ...248 Disposal ...
Page 254: ...254 Contact ...
Page 256: ...256 Recommended Literature ...
Page 266: ...266 Appendix Figure 225 Declaration of conformity ...
Page 268: ...268 Appendix ...
Page 269: ......