181
F(L)CS Data Acquisition
18.1.3
Testing the Positioning Accuracy
In order to test the positioning accuracy of the scanner on different pixels, the multipoint
functionality for FCS can be used for a different purpose.
This test is not necessary to ensure the functional capability of the system as such; you use
it only to confirm the performance capability of the system for yourself.
To carry out this test, proceed as follows:
1. Switch off all external detectors (the APD detector unit or the APDs with the DSN)
electrically.
2. Start the SMD workstation.
3. Start the SymPhoTime software.
4. Create a new working area by selecting
New Workspace
in the
File
menu.
5. Using your bleach specimen, create a well-illuminated xy image as described in
in the
Setup Imaging
step of the FCS Wizard.
Use the photomultipliers for this (
).
Use the zoom that you want to test.
Use the same scan speed as the one you used previously when calibrating the scanner.
6. Select the
Measurements
step in the FCS Wizard (
Observe the SymPhoTime manual
You will find information about creating a new working area in the
SymPhoTime manual.
Summary of Contents for TCS SP8 SMD
Page 1: ...10 Living up to Life User Manual Leica TCS SP8 SMD for FCS FLIM and FLCS ...
Page 4: ...4 Copyright ...
Page 14: ...14 Contents ...
Page 18: ...18 Intended Use ...
Page 20: ...20 Liability and Warranty ...
Page 28: ...28 General Safety Notes ...
Page 32: ...32 Additional Notes on Handling the System ...
Page 44: ...44 System Overview and Properties ...
Page 60: ...60 SMD Components Figure 31 DSN 102 Dual SPAD Power Supply ...
Page 80: ...80 Safety Features ...
Page 102: ...102 Switching On the System ...
Page 116: ...116 LAS AF ...
Page 214: ...214 Changing the Specimen ...
Page 216: ...216 Changing the Objective ...
Page 238: ...238 Switching Off the System ...
Page 242: ...242 Repairs and Service Work ...
Page 244: ...244 Maintenance ...
Page 246: ...246 Disassembly and Transport ...
Page 248: ...248 Disposal ...
Page 254: ...254 Contact ...
Page 256: ...256 Recommended Literature ...
Page 266: ...266 Appendix Figure 225 Declaration of conformity ...
Page 268: ...268 Appendix ...
Page 269: ......