208
F(L)CS Data Acquisition
18.4.6
Operating the FCS Measurement Series
The measurement series is started by clicking
Run FCS
image acquisition button. It stops
automatically after all measurements are taken. A user-defined stop is possible by clicking
Stop FCS
In the count rate monitor (
), the current count rate (in cps – counts per second) of
each channel as well as the counts per molecule (in cpm – counts per second per molecule)
are displayed during the individual measurements.
In SymPhoTime, an online correlation curve is displayed and saved together with the raw
data.
Both the count rate and correlation curve give a first impression about the quality of the
measurements. Images acquired automatically during the FCS measurement series (with
Image after each __ cycles
,
) can indicate movements of the specimen.
Summary of Contents for TCS SP8 SMD
Page 1: ...10 Living up to Life User Manual Leica TCS SP8 SMD for FCS FLIM and FLCS ...
Page 4: ...4 Copyright ...
Page 14: ...14 Contents ...
Page 18: ...18 Intended Use ...
Page 20: ...20 Liability and Warranty ...
Page 28: ...28 General Safety Notes ...
Page 32: ...32 Additional Notes on Handling the System ...
Page 44: ...44 System Overview and Properties ...
Page 60: ...60 SMD Components Figure 31 DSN 102 Dual SPAD Power Supply ...
Page 80: ...80 Safety Features ...
Page 102: ...102 Switching On the System ...
Page 116: ...116 LAS AF ...
Page 214: ...214 Changing the Specimen ...
Page 216: ...216 Changing the Objective ...
Page 238: ...238 Switching Off the System ...
Page 242: ...242 Repairs and Service Work ...
Page 244: ...244 Maintenance ...
Page 246: ...246 Disassembly and Transport ...
Page 248: ...248 Disposal ...
Page 254: ...254 Contact ...
Page 256: ...256 Recommended Literature ...
Page 266: ...266 Appendix Figure 225 Declaration of conformity ...
Page 268: ...268 Appendix ...
Page 269: ......