174
Summarized manual for FLIM experiments
17.12.2
Running IRF Measurement
Carry out the following steps to measure the IRF:
1. In the FLIM Wizard in LAS AF, go to step
Setup FLIM
.
2. Start the scanning operation in test mode by selecting the image acquisition button
Run
FLIM Test
. Normally, the SymPhoTime software will start the data acquisition
automatically. However, you can also start the FLIM data acquisition by selecting the
Acquire LSM Measurements
button.
3. Enable the
Time Trace
tab in the measurement preview window.
4. Adjust the focus of the microscope in order to maximize the displayed count rate.
5. Set the count rate to approximately 50,000 count/s.
6. The laser output power must not be changed on the laser driver itself, but only on the
laser coupling module (LCU) or on the scaffold on the laser head.
Changing the laser driver settings typically changes the shape and temporal position of the
laser pulse. You can use the driver settings to optimize the pulse shape of the laser for
shortest IRF width.
7. You can now test the fading behavior of your specimen in the oscilloscope window in the
SymPhoTime software. For this purpose, quit the preview and select
Show
Oscilloscope...
.
8. Select the
TCSPC Histogram
tab, select
Start
and monitor the fading of the fluorescence
as shown in
. In the oscilloscope window, the time width per channel can also
be set.
Data is not saved.
Data from a scan in test mode is not saved.
Summary of Contents for TCS SP8 SMD
Page 1: ...10 Living up to Life User Manual Leica TCS SP8 SMD for FCS FLIM and FLCS ...
Page 4: ...4 Copyright ...
Page 14: ...14 Contents ...
Page 18: ...18 Intended Use ...
Page 20: ...20 Liability and Warranty ...
Page 28: ...28 General Safety Notes ...
Page 32: ...32 Additional Notes on Handling the System ...
Page 44: ...44 System Overview and Properties ...
Page 60: ...60 SMD Components Figure 31 DSN 102 Dual SPAD Power Supply ...
Page 80: ...80 Safety Features ...
Page 102: ...102 Switching On the System ...
Page 116: ...116 LAS AF ...
Page 214: ...214 Changing the Specimen ...
Page 216: ...216 Changing the Objective ...
Page 238: ...238 Switching Off the System ...
Page 242: ...242 Repairs and Service Work ...
Page 244: ...244 Maintenance ...
Page 246: ...246 Disassembly and Transport ...
Page 248: ...248 Disposal ...
Page 254: ...254 Contact ...
Page 256: ...256 Recommended Literature ...
Page 266: ...266 Appendix Figure 225 Declaration of conformity ...
Page 268: ...268 Appendix ...
Page 269: ......