192
F(L)CS Data Acquisition
Figure 169:
Acquiring the reference image using continuous wave lasers and photomultipliers
18.2.2.2
Using MP Lasers
1. For MP scanning operations, select the desired wavelength.
2. Choose a suitable neutral density filter for the attenuation using the
Trans
slider and/or
adjust the electro-optical modulator (EOM) using the
Gain
and
Offset
sliders.
3. If the system has a multifunction port, it should be set to
SP680
or
SP700
.
For more detailed information about switching on the MP lasers, refer to
.
18.2.2.3
Using Pulsed VIS Lasers
1. Close the shutter of the continuous wave VIS laser in
Beam Path Settings
and move all
acousto-optical tunable filters to the 0 position.
2. Open the shutter for pulsed lasers (
.
3. Activate the desired laser by moving the corresponding slider upwards (
).
Summary of Contents for TCS SP8 SMD
Page 1: ...10 Living up to Life User Manual Leica TCS SP8 SMD for FCS FLIM and FLCS ...
Page 4: ...4 Copyright ...
Page 14: ...14 Contents ...
Page 18: ...18 Intended Use ...
Page 20: ...20 Liability and Warranty ...
Page 28: ...28 General Safety Notes ...
Page 32: ...32 Additional Notes on Handling the System ...
Page 44: ...44 System Overview and Properties ...
Page 60: ...60 SMD Components Figure 31 DSN 102 Dual SPAD Power Supply ...
Page 80: ...80 Safety Features ...
Page 102: ...102 Switching On the System ...
Page 116: ...116 LAS AF ...
Page 214: ...214 Changing the Specimen ...
Page 216: ...216 Changing the Objective ...
Page 238: ...238 Switching Off the System ...
Page 242: ...242 Repairs and Service Work ...
Page 244: ...244 Maintenance ...
Page 246: ...246 Disassembly and Transport ...
Page 248: ...248 Disposal ...
Page 254: ...254 Contact ...
Page 256: ...256 Recommended Literature ...
Page 266: ...266 Appendix Figure 225 Declaration of conformity ...
Page 268: ...268 Appendix ...
Page 269: ......