Ambient Conditions
8.4
Electrical Connection Requirements
WARNING
Electric shock is possible when using ungrounded sockets
This system is designed for connection to grounded (earthed) outlets.
The grounding plug performs an important safety function. To avoid the
risk of electrical shock or damage to the instrument, do not disable this
feature. Operation without grounded sockets is not permitted.
8.4.1
System with Flexible Supply Unit
Supply voltage
100 V~ to 240 V~ ± 10%, grounded
Frequency
50/60 Hz
Power connection
Three separate electric circuits with
Power supply voltage: 100 V - 120 V, fuse protection: 20 A
or:
Power supply voltage: 200 V - 240 V, fuse protection: 12 - 16 A
(Two for the flexible supply unit's power supply and one for
supplying peripheral devices)
To prevent damage to the instrument, the two types of
voltage (200 to 240 V~ and 100 to 120 V~) must not be
combined. Depending on the system configuration, further
electrical circuits may be required.
Power supply for USA
Two terminals of the NEMA 5-20 type for the supply unit
Terminals of the NEMA 5-15 type for the remaining
instruments
for the SMD rack
a separate electric circuit with
Power supply voltage: 100 V~ to 240 V~, fuse protection: 16 -
20 A, dependent on local regulations
for the infrared laser and its
cooling system (MP
configuration)
a separate electric circuit with
Power supply voltage: 100 V~ to 240 V~, fuse protection: 12 -
16 A, dependent on local regulations
When using HyD RLDs
An additional power supply
Summary of Contents for TCS SP8 SMD
Page 1: ...10 Living up to Life User Manual Leica TCS SP8 SMD for FCS FLIM and FLCS ...
Page 4: ...4 Copyright ...
Page 14: ...14 Contents ...
Page 18: ...18 Intended Use ...
Page 20: ...20 Liability and Warranty ...
Page 28: ...28 General Safety Notes ...
Page 32: ...32 Additional Notes on Handling the System ...
Page 44: ...44 System Overview and Properties ...
Page 60: ...60 SMD Components Figure 31 DSN 102 Dual SPAD Power Supply ...
Page 80: ...80 Safety Features ...
Page 102: ...102 Switching On the System ...
Page 116: ...116 LAS AF ...
Page 214: ...214 Changing the Specimen ...
Page 216: ...216 Changing the Objective ...
Page 238: ...238 Switching Off the System ...
Page 242: ...242 Repairs and Service Work ...
Page 244: ...244 Maintenance ...
Page 246: ...246 Disassembly and Transport ...
Page 248: ...248 Disposal ...
Page 254: ...254 Contact ...
Page 256: ...256 Recommended Literature ...
Page 266: ...266 Appendix Figure 225 Declaration of conformity ...
Page 268: ...268 Appendix ...
Page 269: ......