Installation
Step
Mains Wiring
Remark
6.1
Connect mains plug.
If the preinstalled cable can’t be used, mains may be
directly connected to Z1. PE must be connected to the
terminal with the earth symbol. The AC supply wires
must be connected to the terminals labeled 1 and 2.
Voltage between terminal 1 and 2 must be 230V±10%
50/60Hz.
The diameter of the cables must be min 2.5 mm. The litz
wires have to be fixed by means of a cable binder in a
min. distance to the terminals.
Mains supply must be
protected by a 16A fuse and
a continuous ground
connection. Use fuses with
slow tripping characteristic or
high surge current capability,
since the switching power
supplies require high inrush
current.
6.2
Connect the NOVOLAS Basic AT system to the electrical
supply.
Check whether the voltage
agrees with the indication on
the type plate.