User Manual for PuraLev
®
600SU
www.levitronix.com
PL-4032-00, Rev06, DCO# 20-144
27
2) Motors used with cooling module ACM-600.3
•
Mount the cooling module
ACM-600.3
to the motor according to
•
Remove one of the four M5 screws on the mounted cooling module.
•
Use a crimp cable lug to connect the earth wire.
•
Attach the crimp cable lug with a M5 x 30 mm stainless steel screw, a washer disc and a spring
lock washer to the motor through the cooling module Figure 32.
Figure 32: Attachment of a protective earth wire to the backside of the motor through ACM-600.3
One screw to be
removed
Parts for earth wire
attachment