User Manual for BPS-200
www.levitronix.com
PL-4012-00, Rev07, DCO# 20-144
7
2.2.3 ATEX / IECEx System Configuration
Together with the standard pump-head and controllers an
ATEX / IECEx
certified motor allows installation of
motor and pump-head within an Ex classified area (see
). The
ATEX / IECEx
motor (
Pos. 2c
in
) comes delivered with special connectors and according extension cables (see
). One option to
lead the cables outside of the Ex area is an Ex certified cable sealing system as listed in
Pos.
8
).
Pump Head
Fluid In
Fluid Out
Motor Power Cable
Adaptor/Extension Cable
Power Supply
24-48 VDC / 200 W
All
Controller
Types
Ex conform cable sealing system
Cabinet Boundary for
Ex Classified Area
Ex Classified Location
Chemical Resistance and IP-67 Space
Motor
(ATEX/IECEx)
Figure 5: System configuration for ATEX / IECEx applications
2.3 General Environmental Conditions
Usage
Indoor
1
Altitude
Up to 2000
m
Operating ambient temperature
0 to 40
0
C
Storage ambient temperature
(Extremes for Transportation)
-20 to 80
0
C
Operating humidity range (relative humidity)
15
– 95
%
(non-condensing)
Storage humidity range (relative humidity)
(Extremes for Transportation)
15
– 95
%
(non-condensing)
Normal storage conditions
Ambient temp.: 20 to 30
0
C
Relative humidity: 50
%
(non-condensing)
DC supply fluctuations
10% of nominal voltage
Transient over-voltages typically present on
the mains supply
Surge immunity according to EN 61000-4-5
(tested together with certified AC/DC power supply)
Pollution degree
2
Table 5: Environmental conditions for pump system
Note 1: Contact Levitronix
®
technical department (see Section 8) for outdoor usage.
1
2c
4
3b
8