User Manual for BPS-2000
www.levitronix.com
PL-4021-00, Rev06, DCO# 21-037
32
4.2 Mechanical Installation of the Pump/Motor
4.2.1 Standard Installation Instructions and Information
!
WARNING
Overheating of the Motor Power and Extension Power Cable
To prevent an overheating of the motor power and extension power
cables, do not roll-up or install several motor power cables in the same
cable channel. This is has especially to be considered when long motor
power cables are used.
▪
The motor can be fixed with four screws on the motor feet (see
).
▪
As an alternative the motor can be mounted with 12 screws on the back (see
).
▪
The motor can be mounted in either the horizontal or vertical position.
▪
Minimum cable bending radius according
to shall be followed for the installation of the
motor and extension/adaptor cables.
▪
Each motor is identified with a unique serial number. This serial number consists of a series of
6 digits were the 5th and the 6th digit representing the manufacturing year.
▪
To prevent an overheating of >90 °C of the motor power cable in extension power cable,
please don’t roll-up or install several motor power cables in the same cable channel.
4.2.2 Installation of Ex and Hazardous Location Certified Motors
!
WARNING
Motors for ATEX / IECEx and Hazardous Location Cl1 Div2
applications. Only specific types of motors LPM-2000 are classified for
the use in Ex classified locations. Refer to the corresponding section in
the manual.
!
WARNING
Motors for ATEX / IECEx and Hazardous Location Cl1 Div2
applications. Use only, if necessary, the cooling module ACM-4.3 for
motors installed in Ex classified locations. The use of other materials
for the air cooling module is not allowed.
!
WARNING
Motors for ATEX / IECEx applications. Cable and cable glands of the
motor must be protected against impact energy.
!
WARNING
For liquid temperatures above 75 °C in Ex applications active cooling
using the cooling module ACM-4.3 is needed.