User Manual for BPS-iF100
www.levitronix.com
PL-4037-00, Rev02, DCO# 21-057
9
2.4 Pressure-Flow Curves
Figure 8: Pressure/flow rate curves
(Measured with pump head LPP-200.7)
2.5 Maximum Static Pressure for Pump Heads
Figure 9: Specification for maximum static pressure of pump head LPP-200.7
(Pressure limits only valid for pump head mounted on motor)
0.0
1.0
2.0
3.0
4.0
5.0
0
5
10
15
20
25
30
0.00
0.25
0.50
0.75
1.00
1.25
1.50
1.75
2.00
2.25
0.0
2.5
5.0
7.5
10.0
12.5
15.0
17.5
20.0
[gallons/min]
[psi]
[bar]
[liters/min]
9000 rpm
4000 rpm
Specific gravity = 1 g/cm
3
Viscosity = 0.7 cP
Liquid Temp.: 40
°C
8000 rpm
7000 rpm
6000 rpm
5000 rpm
0
20
40
60
80
100
77
87
97
107
117
127
137
147
157
167
177
187
0
1
2
3
4
5
6
7
8
25
30
35
40
45
50
55
60
65
70
75
80
85
90
[psi]
[°F]
[bar]
Liquid Temperature in [°C]
Max. Operational Pressure for LPP-200.7
Max. Static Pressure for LPP-200.7 (PFA Housing, PP Reinforcing Cup with SS Screws)
Safety Factor > 1.3 (according to EN12162)
Note: Burst pressure is significantly higher.
Ambient Temp = 25
°C
Note: Pressure limit only valid for pump head mounted on motor.