User Manual for PuraLev
®
2000SU
www.levitronix.com
PL-4034-00, Rev02, DCO# 20-144
26
4.2 Mechanical Installation of the Motor
4.2.1 Standard Installation Instructions and Information
!
WARNING
Overheating of the Motor Power and Extension Power Cable
To prevent an overheating of the motor power and extension power
cables, do not roll-up or install several motor power cables in the same
cable channel. This is has especially to be considered when long motor
power cables are used.
▪
The motor can be fixed with four screws on the motor feet (see
).
▪
As an alternative the motor can be mounted with 12 screws on the back (see
▪
The motor can be mounted in either the horizontal or vertical position. When mounting in
horizontal position it shall be mounted elevated from the bottom for at least 15mm in order to
assure sufficient space for the pump head socket and for the handling of the mounting screw
of the pump head (see
▪
Minimum cable bending radius according
to shall be followed for the installation of the
motor and extension/adaptor cables
▪
Each motor is identified with a unique serial number. This serial number consists of a series of
6 digits were the 5th and the 6th digit representing the manufacturing year.
▪
To prevent an overheating of > 90°C of the motor power cable in extension power cable,
please don’t roll-up or install several motor power cables in the same cable channel.
Figure 29: Elevated mounting of motor
> 15mm
Free space for mounting screw
and pump head socket.