User Manual for PuraLev
®
600MU
www.levitronix.com
PL-4033-00, Rev09, DCO# 20-144
21
Figure 25: Overview of the controller LPC-600.2 for extended operation
Interface (as labeled)
Description
1
“SENSORIC”
Position, field and temperature sensor signals from motor.
Torque spec. for tightening of connector screws: Min. = 0.4, Max. = 0.6 Nm
2
“USER INTERFACE”
2 Analog Input
- Analog current input: 4
– 20 mA
- 450 Ohm shunt input
2 Analog Input
- Analog voltage input 0
– 10 V
- Direct connection, no galvanic isolation
- 7.8 k
input resistance
2 Analog Output
- Analog voltage output: 0
– 5 V
- Direct connection, no galvanic isolation
- Max. Output current: 2mA
4 Digital Input
- Galvanic isolation with optocoupler
- Lowest input voltage for high level detection: min. 5 V
Typical 24 V / 16 mA, maximal 30 V / 20 mA
- Highest input voltage for low level detection: max. 0.8 V
- Minimum input resistance: R
IN
= 2.2 k
4 Digital Output
- Galvanic isolation with relay
- Relay: 1A / 30VDC, 0.3A / 125 VAC
3
“USB”
USB interface
4
“POWER OUTPUT”
Drive and bearing currents of the motor.
1
Torque spec. for tightening of connector screws on motor side: Min. = 0.5 Nm, Max. = 0.6 Nm
5
“POWER INPUT”
DC power input -> 48V.
1
Torque spec. for tightening of connector screws on motor side: Min. = 0.5 Nm, Max. = 0.6 Nm
6
“Power on” Green LED
LED is on if supply voltage of signal electronics is present.
7
“Power Output not active”
Red LED
LED is off if the switched output stage of the controller is enabled. If the LED is on, the bearing
and drive coils of the motor carry no current.
8
“RESET” Button
Reset button of the controller stage
Table 10: Description of interfaces of LPC-600.2 controller
1: Connectors are not made for multiple connection cycles. Avoid connections cycles > 25.
1
2
3
4
5
7
8
6