User Manual for PuraLev
®
i100SU
www.levitronix.com
PL-4063-00, Rev03, DCO# 20-275
10
2.4 Pressure-Flow Curves
Figure 11: Pressure/flow rate curves of DCP-200 pump head
2.5 Maximum Static Pressure for Pump Heads
Figure 12: Specification for maximum static pressure of pump head DCP-200
Note 1: Pressure limits only valid for pump head mounted on motor and do not account for tubing limits attached to the fittings.
Note 2: The above pressure chart also applies to the pump head DCP-200.3-SF1-G25 with sterile fittings.
0
5
10
15
20
25
30
0.0
1.0
2.0
3.0
4.0
5.0
0.00
0.25
0.50
0.75
1.00
1.25
1.50
1.75
2.00
2.25
0.0
2.5
5.0
7.5
10.0
12.5
15.0
17.5
20.0
[psi]
[gallons/min]
[bar]
[liters/min]
Specific gravity = 1 g/cm
3
Viscosity = 0.7 cP
Liquid Temp.: 40
C
DC-Voltage = 24VDC
Pumphead: DCP-200.2
8000 rpm
7000 rpm
6000 rpm
5000 rpm
4000 rpm
9000 rpm
0
10
20
30
40
50
60
70
80
77
87
97
107
117
127
137
0
1
2
3
4
5
6
25
30
35
40
45
50
55
60
[psi]
[°F]
[bar]
Liquid Temperature in [°C]
Max. Static Pressure for DCP-200
Safety Factor = 1.3 (according to EN12162)
Limitation: Given by deformation. Burst pressure is significantly higher
Ambient Temp = 25
C
Note: Pressure limit only valid for pumphead mounted on motor. Values
only valid after Gamma radiation doses of < 40 kGy.