ADVANCED MICRO CALIBRATION BATH TEMPERATURE
LCA 30/50
www.leyroinstruments.com
22
Second:
Press
Add Step
or press
Actions
and
Insert
. You must add another step called
Wait For
.
For it, press
Instant Change
and press
Wait For
.
Once this is done, modify the value of
Wait For Process 1 Condition
and
Wait For
Process 2 Condition
.
This step executes an undefined timer, which will end when the set point has been
reached.
ATTENTION:
It should be selected, in
Wait for Process 1 Condition
and
Wait for
Process 2 Condition
,
Above
when the temperature are heating and
Below
when
the temperature are cooling.
Then, press
Done
.
If you have modified the value, but want to change it, press
Actions
and
Edit Step
Parameters
.
If you want to delete a step, click on
Actions
and
Delete Step
.
Third:
Press
Add Step
or press
Actions
and
Insert
. You must add another step called
Soak
.
For it, press
Wait For
and press
Soak
.
Once the set point has been reached, the equipment will be at that point for the time
specified (advisable to modify the value to
0 hours
,
20 minutes
and
0 seconds
).
Then, press
Done
.
If you have modified the value, but want to change it, press
Actions
and
Edit Step
Parameters
.
If you want to delete a step, click on
Actions
and
Delete Step
.
These steps must be repeated for the necessary calibration points, maintaining the order
of the set point, wait and constant value.
In case of not wanting this profile, press cancel new profile and in actions of that same
profile press delete profile.
Example:
Lower the temperature from 25,000ºC to 20,000ºC.
If the temperature is 25,000ºC and we need to lower it to 20,000ºC, the
Instant Change
should be at
20,000ºC
in the
Target Set Point Loop 1
and
Target Set Point Loop 2
The
Profile
is started.
Once the
Instant Change
is started, the equipment will go to the next step,
which will be
Wait For
, which should be configured in
Below
in
Wait For
Process 1 Condition
and
Wait For Process 2 Condition
, since the
temperature is going down.
Once the temperature has been reached and the equipment drops below
20,000ºC, it will go to the last step, which will be
Soak
, which will be
configured in
20 minutes
.
When that time passes, the profile will be finished.
Summary of Contents for LCA 30
Page 6: ...ADVANCED MICRO CALIBRATION BATH TEMPERATURE LCA 30 50 www leyroinstruments com 4 ...
Page 30: ...ADVANCED MICRO CALIBRATION BATH TEMPERATURE LCA 30 50 www leyroinstruments com 28 ...
Page 32: ...ADVANCED MICRO CALIBRATION BATH TEMPERATURE LCA 30 50 www leyroinstruments com 30 ...
Page 34: ...ADVANCED MICRO CALIBRATION BATH TEMPERATURE LCA 30 50 www leyroinstruments com 32 ...
Page 36: ...ADVANCED MICRO CALIBRATION BATH TEMPERATURE LCA 30 50 www leyroinstruments com 34 ...