STEPS
Here the step size for the STEP function, in x % of the set range, can be changed. Possi
-
ble settings are from 1 … 50 %.
UNIT
Here the unit, in which the low-pressure controller will display the measured value on the
screen, can be selected. Dependent upon the measuring range of the instrument, various
units are not available since they could not be displayed.
MODE
Here the operating mode of the controller can be set. The following modes can be selec
-
ted:
Measure (MEAS): The controller measures the pressure in the set pressure unit
Control (CTRL): A set point is automatically regulated
Automatic (AUTO): A stored test routine will be carried out
The menu points
With the MENU button and with the / button, press the up and down until one is at
the required menu point. With the SELECT button, one reaches the desired submenu.
The following items are available in the submenu:
RANGE
Under the RANGE menu point the controller‘s final control value can be changed for the
control.
Navigation within the menu:
OK: Acts to confirm the relevant setting
Left/Right arrow: Setting of the active decimal point
Up/Down arrow: Change between the different submenu points