9. DISMOUNTING, RETURN AND DIPOSAL
Warning
!
Residual media on the dismounted low-pressure controller can result in a
risk to persons, the environment and equipment. Take sufficient precautio
-
nary measures.
Dismounting
Only disconnect test and calibration installations once the system has been depressuri
-
sed!
When disassembling the instrument the following must be followed:
1. Ensure that there is no positive or negative gauge pressure in the instrument and that
all parts of the instrument are at room temperature.
2. Switch the instrument off using the ON/OFF switch
3. First pull the power supply unit from the mains socket and then from the power
supply
socket on the instrument.
4. Loosen the pressure connections
5. Do not close the pressure inputs!
6. Make sure there is no media left in the instrument.
Warning
!
Strictly observe the following when shipping the instrument:
All instruments delivered to Leyro must be free from any kind of hazardous
substances (acids, bases, solutions etc.).
When returning the instrument, use the original packaging or a suitable transport
package.
To avoid damage:
1. Do not return the instrument with the pressure connections closed! Barometric
pressure changes can damage instruments with low pressure ranges.
2. Wrap the instrument in an antistatic plastic film.
3. Place the instrument, along with the shock-absorbent material, in the packaging.
Place shock-absorbent material evenly on all sides of the transport packaging.
4. If possible, place a bag, containing a desiccant, inside the packaging.
5. Label the shipment as transport of a highly sensitive measuring instrument.
Return
Summary of Contents for LPG 300
Page 1: ...LPG 300 PORTABLE LOW PRESSURE CONTROLLER MANUAL ...
Page 2: ......
Page 4: ......
Page 10: ...Electrical connections in the rear USB Port Port PC RS 23 Connection for power supply ...
Page 26: ......
Page 27: ......
Page 33: ......
Page 34: ...FRONT VIEW ...
Page 35: ...SIDE VIEW ...
Page 38: ......