Chapter 1: Pearl Docking Station

6) Connect one of the provided HEPA filter assemblies between the SmartFlow system and the

Induction Chamber.

7) Connect the opposite end of the HEPA filter assembly to the provided anesthesia gas supply tube

and then connect the other end of the anesthesia gas supply tube to one of the output manifold
connectors on the SmartFlow Anesthesia System, or equivalent (Figure 1-5).

Note:

For imaging beds not requiring a HEPA filter, connect the anesthesia gas supply tube dir-

ectly to the rotameter. For other vaporizers, it may be necessary to remove the connector that con-
nects to the SmartFlow System and replace it with an appropriate fitting.

CAUTION:

The Pearl

®

 Imaging System Docking Station is compatible only with anesthesia

systems that supply isoflurane gas. The flow rate and pressure into the instrument should not
exceed 1 liter per minute and 1 psi, respectively. If the pressure exceeds 5 psi, the rotameter
will not accurately show the flow to the nose cone.

ATTENTION:

La station d'amarrage de Pearl n'est compatible qu'avec les systèmes d'anesthésie qui
fournissent le gaz isoflurane. Le débit et la pression dans l'instrument ne devraient pas
excéder 1 litre par minute et 6.9 kPa, respectivement. Si la pression excède l'estimation de
34.5 kPa, le rotamètre ne montrera pas en juste proportion l'écoulement au cône de nez.

D

K

E

F

N

J

K

M

G

J

K

I

A

A

B

C

D

E

F

G

H

I

J

K

L

M

N

O

B

O

C

J

EQUIPMENT LEGEND

EX.

EX.

OXYGEN TANK

OXYGEN REGULATOR

SMARTFLOW ANESTHESIA SYSTEM

SMARTFLOW VAPORIZER

INDUCTION CHAMBER WITH HEATER

WATER PUMP

PEARL IMAGER

PEARL ROTAMETER

PEARL IMAGER IMAGING BED OR CLEAN BOX

CHARCOAL FILTER CANISTER

HEPA FILTER ASSENBLY (9957-055) 
FOR 9930-022 PEARL CLEAN BOX

DOCKING STATION ROTAMETER AND STAND

PEARL DOCKING STATION

10’ TUBING ASSEMBLY (#9957-056) INCLUDING
 WITH THE PEARL DOCKING STATION

TUBING ASSEMBLY INCLUDDED WITH 
THE PEARL IMAGER

OUT

OUT

OUT

IN

IN

IN

H

L

Figure 1-5. General representation of a typical system consisting of the docking station with the Pearl Imaging System
and SmartFlow Anesthesia System. Docking station rotameter (L) connects to optional HEPA filter assembly (K) and
anesthesia gas supply tube (N) before connecting to a port on the SmartFlow Anesthesia System (C).

Page 4

Summary of Contents for 5700-DS

Page 1: ...Pearl Trilogy Accessories Manual...

Page 2: ......

Page 3: ...This product model number 5700 DS is a CE marked product For conformity information contact LI COR Support at http www licor com biotechsupport Outside of the U S contact your local sales office or d...

Page 4: ...nt NOTE Notes contain additional information and useful tips IMPORTANT Information of importance to prevent procedural mistakes in the operation of the equipment or related software Failure to comply...

Page 5: ...d by copyright All rights are reserved No part of this document may be photocopied reproduced or translated to another language without prior written consent of LI COR Inc Printing History Publication...

Page 6: ...Box Description 9 Installation 9 Connecting a HEPA Filter to the Docking Station 9 Connecting a HEPA Filter to the Pearl Imager 10 Nose Cone Installation and Replacement 10 Sanitizing the Pearl Clean...

Page 7: ...on trol to a Pearl accessory imaging bed outside of the Pearl Imager allowing animals to be prepared for imaging at an external station l Warms the heater plate in any of the Pearl accessory imaging b...

Page 8: ...instructions dans ce chapitre pour vous assurer que le port d chappement est reli un filtre charbon que le syst me d anesth sie fourni par l utilisateur est correctement reli et que le cone de nez est...

Page 9: ...UST on the rotameter to the port labeled OUT on the back of the docking station 5 If using the Pearl Clean Box connect one of the provided HEPA filter assemblies packaged with the Clean Box to the por...

Page 10: ...rage de Pearl n est compatible qu avec les syst mes d anesth sie qui fournissent le gaz isoflurane Le d bit et la pression dans l instrument ne devraient pas exc der 1 litre par minute et 6 9 kPa resp...

Page 11: ...UST port Recommendations for monitoring and replacement of the charcoal filter are given in Chapter 3 of the Pearl Imager Operator s Manual Figure 1 6 The included charcoal filter or a user supplied c...

Page 12: ...structions for the specific imaging bed you are using WARNING Make sure a black nose cone plug is tightly inserted in the nose cone before turning on the anesthesia gas Failure to insert a nose cone p...

Page 13: ...the adjustment knob on the rotameter to set the flow rate to 0 5 liters per minute typical and start gas flow from isoflurane vaporizer or equivalent Allow 1 minute to prime the system with isoflurane...

Page 14: ...ire or loose connections Periodically inspect external tubing for cracking or damage that could cause leaks Docking Station Specifications Docking Station Operating Conditions For Indoor use only oper...

Page 15: ...the anesthesia gas When the Clean Box is undocked the gas ports automatically close The Pearl Clean Box window is a specially coated cell cast acrylic window for optimum light transmission and durabi...

Page 16: ...ion after it is plugged in 3 Connect the opposite end of the HEPA filter assembly to the provided anesthesia supply tube provided with the Pearl Imager Nose Cone Installation and Replacement Nose cone...

Page 17: ...the imaging bed surface with 0 5 chlorine bleach Ethanol or isopropanol can be also used as a cleaning solution for applications where chlorine bleach is not required If necessary clean any fingerprin...

Page 18: ...IMPORTANT Move the Clean Box quickly Anesthesia gas does not flow while the Clean Box is undocked 8 Remove the Clean Box cover to expose the window close the imaging drawer and acquire an image as des...

Page 19: ...rove insufficient use LI COR Optical Tissue and LI COR Optical Cleaner Apply a few drops to an optical tissue and gently clean the window with a circular motion Follow with a second dry tissue to remo...

Page 20: ...iosciences GmbH Siemensstra e 25A 61352 Bad Homburg Germany Phone 49 0 6172 17 17 771 Fax 49 0 6172 17 17 799 bio eu licor com LI COR Biosciences UK Ltd St John s Innovation Centre Cowley Road Cambrid...

Reviews: