Section 4
4-2
Calibration
■
To set the H
2
O Span
1.
Setup the LI-610 for an appropriate dew point
Subtract about 5 °C from room temperature,and use that for the target
dew point temperature. Wait until the condensor’s temperature (as
monitored on the LI-610) reaches this target. The reason for this 5
°C “buffer” is to avoid condensation in the line between the LI-610’s
condensor and the IRGA. If condensation happens, you will have
large errors.
2.
Set the flow rate
Use a flow rate of about 0.5 l min
-1
from the LI-610.
3.
Attach to the IRGA
You have two choices, as shown in Figure 4-1. We recommend
option B: splitting the flow and connecting both the reference and
sample, with Match Off. The reason for this is that you will be able
to drastically reduce the equilibrium time, waiting for the sample cell.