Troubleshooting
Light Source / Sensor Problems
Using the LI-6400
20-31
2
Blown fuse?
The flow board fuse protects the lamp. If it is blown, a number of things will
not work, including the lamp, lamp fan, pump and cooling fans.
3
Check the lamp voltage
Unscrew the lamp connector (Figure 20-7) to gain access to the wires in the
connector. Measure the voltage (with the lamp turned on high) between the
orange and white wires.
Be careful, there should be voltage in excess of 100
Volts.
If the voltage is near 0: A cable or connector problem is indicated. Continue
with Step 4.
If the voltage is about 12V: The switching power supply (itÕs in the console)
for the lamp has failed. Contact LI-COR.
If the voltage is over 100V: (Normal)The problem may be a broken connec-
tion within the light source itself. Contact LI-COR.
4
Check the 26 pin D connector
Make sure none of the pins have been pushed in or broken (Table 20-1 on
page 20-45).
5
Check the cable
Try a different cable, if one is available to you.
Figure 20-7. Unscrew the lamp connector to access the four wires.
Caution -
High Voltage.
Summary of Contents for LI-6400
Page 1: ...Using the LI 6400 Portable Photosynthesis System ...
Page 15: ...Part I The Basics ...
Page 16: ......
Page 174: ...Making Measurements Answers to Questions 4 56 Using the LI 6400 4 ...
Page 175: ...Part II Useful Details ...
Page 176: ......
Page 200: ...Standard Tools Power ON Hooks 5 24 Using the LI 6400 5 ...
Page 214: ...Real Time Data Real Time Graphics 6 14 Using the LI 6400 6 ...
Page 234: ...Environmental Control Light Control 7 20 Using the LI 6400 7 ...
Page 244: ...Light Sensor Considerations Gallium Arsenide Phosphide GaAsP Sensor 8 10 Using the LI 6400 8 ...
Page 288: ...Data Logging Making Your Own AutoPrograms 9 44 Using the LI 6400 9 ...
Page 289: ...Part III Working With Files ...
Page 290: ......
Page 312: ...The LPL File System Troubleshooting 10 22 Using the LI 6400 10 ...
Page 340: ...Downloading Files Using a Data Capture Program 11 28 Using the LI 6400 11 ...
Page 375: ...Part IV Configuration Issues ...
Page 376: ......
Page 420: ...Defining User Variables Old Style vs New Style 15 18 Using the LI 6400 15 ...
Page 454: ...Using an Energy Balance Further Reading 17 12 Using the LI 6400 17 ...
Page 455: ...Part V Maintenance Troubleshooting ...
Page 456: ......
Page 572: ...Troubleshooting Useful Information 20 46 Using the LI 6400 20 ...
Page 593: ...Part VI Programming ...
Page 594: ......
Page 622: ...Programming with LPL Compiler Directives 22 28 Using the LI 6400 22 ...
Page 846: ...Index I 16 Using the LI 6400 ...