background image

Introduction

The LI-8150 is a robust and simple to use multiplexer
designed to interface up to sixteen soil CO

2

 flux cham-

bers with the LI-8100A Automated Soil CO

2

 Flux

System. Here we describe considerations for construct-
ing a multiplexed system for measuring CO

2

 evolution

in flasks utilizing the LI-8100A and LI-8150. We com-
plete this example by demonstrating a twelve flask
system used to measure respiration of detached fruit.

Flask Volume, Mixing and Flow
Considerations

Volume

The LI-8100A operates as a closed system, determining
CO

2

 flux from the change in CO

2

 mole fraction over

time in a fixed volume. In its simplest form (ignoring
dilution due to water vapor and not standardizing the
rate for sample area or mass) the relationship between
carbon flux and the change in CO

2

 mole fraction with

time takes the following form

F

PV

RT

dC

dt

=

(1)

where F is the flux in 

μ

mol s

-1

, P is pressure in kPa, V is

the system volume in liters, R is the ideal gas constant, T
is temperature in K, and dC/dt is the change in CO

2

mole fraction with time (

μ

mol mol

-1

s

-1

). At ambient

pressure, system volume is the single largest factor
controlling the relationship between dC/dt and flux.
Given the same flux, changing the gas temperature from
20 to 40 

°

C will only change dC/dt by about 7%, while

doubling the system volume will cut dC/dt nearly in half
because the increased volume serves to dilute the change
in CO

2

 mole fraction within the system. As dC/dt

decreases, the potential for error in the measurement

due to instrument noise increases. For the LI-8100A this
noise is known and its effects have been modeled over a
range of dC/dt for various sampling periods (Figure 1).

A chamber volume should be chosen that over the range
of expected fluxes, ensures dC/dt is large enough to be
outside the influence of instrument noise. In circum-
stances where there are constraints on chamber size, the
observation length used by the LI-8100A can be adjusted
to improve the estimate of dC/dt, as increasing the
number of data points used in the calculation will de-
crease the effect of noise (Figure 1).

Figure 1.

 The relationship between potential measurement

error due to instrument noise and the change in CO

2

 mole

fraction with time. Lines are fit to data modeled over 60
(open boxes), 120 (closed boxes), 180 (open circles) and
300 (closed circles) second sampling periods (i.e.,
observation length minus the deadband).

Figure 2 shows the relationship between dC/dt and the
flux for several chamber volumes at standard temperature
and pressure. This, and the information in Figure 1, can
be used as guides for selecting an appropriate chamber
volume and observation length when designing an
experiment.

dC/dt (µmol mol

-1

 s

-1

)

Fractional Error

0.001

0.01

0.1

1

1.000

0.100

0.010

0.001

0.000

Measurement of CO

2

 Evolution

in a Multiplexed Flask System

Continued on next page

App. Note 127

Summary of Contents for LI-8100A

Page 1: ...early in half because the increased volume serves to dilute the change in CO2 mole fraction within the system As dC dt decreases the potential for error in the measurement due to instrument noise incr...

Page 2: ...s the mechanism by which CO2 and water vapor are exported from the system between measurements preventing condensation and excessively high CO2 mole fractions from developing in the flask The flushing...

Page 3: ...o brass plugs from the right hand upper and lower solenoid manifolds 2 Using an adjustable wrench replace the three brass plugs with quick connect elbows part 300 07474 and the one steel plug with a q...

Page 4: ...for flask measurements bottom right The flow path used during measurements is shown with blue arrows and the flow path used to flush flasks between measurements is shown with red arrows In both diagra...

Page 5: ...ver plate from the instrument by loosening the two thumb screws Figure 7A Mount the pump in the LI 8150 case by attaching the mounting plate to one of the two threaded holes normally used to attach th...

Page 6: ...thod 2 If dedicating a port to vent chamber exhaust and supply air to the auxiliary pump is undesirable the supply exhaust lines can be routed out of the LI 8150 by leaving the case open If using this...

Page 7: ...ments are made at near ambient tempera tures a measurement of chamber air temperature may not be required Figure 9 Mason jar lid fitted with plumbing connections and a soil temperature thermistor Two...

Page 8: ...as measured with a thermistor connected to V2 A volume correction can be made for the sample volume by entering a negative value for Extension Tube Volume cm3 Soil CO2 Flux System and LI 8150 Multiple...

Page 9: ...on air trapped in the measurement loop from the previous measurement with the air in the flask Figure 12 The change in CO2 mole fraction with time calculated for various deadband lengths using the dat...

Page 10: ...lated soil samples Jones and Kielland 2002 Haney et al 2008 and monitoring carbon fluxes from attached leaves and fruit Araki et al 1998 Kitano et al 1997 References Araki T M Kitano M Hamakoga and H...

Page 11: ...8100A 8150 770 01 8150 661 8150 203 8150 670 16 or 8150 670 8 196 10534 210 01958 1 9981 173 9981 177 9881 204 140 04315 8150 250 300 07124 300 07125 300 07126 300 07127 167 07256 300 08118 300 07474...

Page 12: ...licor com envsupport UK licor com LI COR Biosciences Global Headquarters 4647 Superior Street Lincoln Nebraska 68504 Phone 1 402 467 3576 Toll free 800 447 3576 Fax 1 402 467 2819 envsales licor com...

Reviews: