background image

manifold, the hose connecting the lower left solenoid
manifold to the pump will have to be temporarily
removed.  First, remove the two thumb screws that
secure the control panel to the LI-8150 and the
ribbon cable from the side of the control panel, and
set the control panel aside (Figure 

4

B).  Carefully

remove the hose from the quick-connect fitting by
pressing down on the orange ring and pulling out on
the hose. Rotate the fitting out of the way.  After the
second quick-connect fitting has been added replace
the hose, seating it firmly in the fitting, and reinstall
the control panel (Figure 

4

C).

3. Ventilation to the flasks is provided with an 8 L/min

diaphragm pump (part #9981-173) that can be
installed inside the LI-8150.  If installing this pump
in a sixteen port multiplexer remove the quick-
connect straight union installed on the pump outlet
and replace with a quick-connect “Y” fitting (part
#300-03367) as shown in Figure 6B. If installing in an
eight port multiplexer no modification is needed
(Figure 6A).

6. Locate the main power supply cable inside the

LI-8150 (part #9981-108; Figure 8). One end of this
cable terminates in a four-pin slip connector con-
nected to the main circuit board and the other in a
four-pin Eurofast bulkhead mounted to the side of
the LI-8150 case. Remove the cable by disconnecting
the four pin slip connector from the circuit board
and removing the nut on the outside of the LI-8150
from the Eurofast bulkhead.

4. Before installing the pump, attach the pump assem-

bly power cable (part #9981-177). This cable con-
nects to the pump via a three-pin slip connector on
the underside of the pump.

5. Remove the cover plate from the instrument by

loosening the two thumb screws (Figure 7A). Mount
the pump in the LI-8150 case by attaching the
mounting plate to one of the two threaded holes
normally used to attach the cover plate (Figure 7B).
Use one of the three 3/8 inch #6-32 hex cap screws
(part #140-04315) and the 7/64 inch long arm hex
key (provided) to attach the mounting plate to the
LI-8150. Use the remaining two hex cap screws to
attach the pump to the plate.

Figure 7. 

The pump mounting plate can be installed inside

the LI-8150 case using one of the two threaded holes
normally used to attach the cover plate. 

A.

 Loosen the two

thumb screws (indicated with arrows) to remove the cover
plate. 

B. 

Use one 3/8 inch long #6-32 hex cap screw to

install the mounting plate in place of the cover. 

C.

 Use the

remaining two screws to attach the pump to the mounting
plate.

7. Install the pump assembly power cable into the

multiplexer in place of the power cable removed in
step 6 (Figure 8). When the new Eurofast bulkhead is
installed the nut on the bulkhead needs to be tight-
ened to 72 inch-lbs of torque to ensure a water tight
seal.

8. Cut two (one for an eight port multiplexer) 48 cm

sections of Bev-a-line tubing.  The two pieces of
tubing are kept at the same length to help ensure that
flow from the pump is split equally between the two
manifolds.  For a 16-port multiplexer insert one end
of each hose into the “Y” fitting (part #300-03367)
installed in step 3, and insert the other end into one
of the quick-connect fittings added to the lower
solenoid manifolds in step 2.

Figure 6.

  

A.

 No modifications are needed to the pump if

installing it in an eight port multiplexer. 

B. 

This pump is

being installed in a sixteen port multiplexer, so a quick
connect “Y” union is installed on the outlet, in place of the
straight union the pump comes with.

A.

B.

A.

5

B.

C.

Summary of Contents for LI-8100A

Page 1: ...early in half because the increased volume serves to dilute the change in CO2 mole fraction within the system As dC dt decreases the potential for error in the measurement due to instrument noise incr...

Page 2: ...s the mechanism by which CO2 and water vapor are exported from the system between measurements preventing condensation and excessively high CO2 mole fractions from developing in the flask The flushing...

Page 3: ...o brass plugs from the right hand upper and lower solenoid manifolds 2 Using an adjustable wrench replace the three brass plugs with quick connect elbows part 300 07474 and the one steel plug with a q...

Page 4: ...for flask measurements bottom right The flow path used during measurements is shown with blue arrows and the flow path used to flush flasks between measurements is shown with red arrows In both diagra...

Page 5: ...ver plate from the instrument by loosening the two thumb screws Figure 7A Mount the pump in the LI 8150 case by attaching the mounting plate to one of the two threaded holes normally used to attach th...

Page 6: ...thod 2 If dedicating a port to vent chamber exhaust and supply air to the auxiliary pump is undesirable the supply exhaust lines can be routed out of the LI 8150 by leaving the case open If using this...

Page 7: ...ments are made at near ambient tempera tures a measurement of chamber air temperature may not be required Figure 9 Mason jar lid fitted with plumbing connections and a soil temperature thermistor Two...

Page 8: ...as measured with a thermistor connected to V2 A volume correction can be made for the sample volume by entering a negative value for Extension Tube Volume cm3 Soil CO2 Flux System and LI 8150 Multiple...

Page 9: ...on air trapped in the measurement loop from the previous measurement with the air in the flask Figure 12 The change in CO2 mole fraction with time calculated for various deadband lengths using the dat...

Page 10: ...lated soil samples Jones and Kielland 2002 Haney et al 2008 and monitoring carbon fluxes from attached leaves and fruit Araki et al 1998 Kitano et al 1997 References Araki T M Kitano M Hamakoga and H...

Page 11: ...8100A 8150 770 01 8150 661 8150 203 8150 670 16 or 8150 670 8 196 10534 210 01958 1 9981 173 9981 177 9881 204 140 04315 8150 250 300 07124 300 07125 300 07126 300 07127 167 07256 300 08118 300 07474...

Page 12: ...licor com envsupport UK licor com LI COR Biosciences Global Headquarters 4647 Superior Street Lincoln Nebraska 68504 Phone 1 402 467 3576 Toll free 800 447 3576 Fax 1 402 467 2819 envsales licor com...

Reviews: