background image

Removing the Regular Resusci

®

 

Junior* Chest Skin

In order to install our Child Defibrillation 

Chest Skin, you must first remove the 

outer skin that is currently in place on 

your CPR manikin.
1. Unhook the chest skin from the 

  sides and fold it up over the head.
2. Lift out the lung plate.
3. Remove the lung bag system from 

  the Resusci

®

 Junior* body.

Figure 3

4. Lift out the chest plate and remove 

  the roll of bags. 

(See figure 3.)

Note:

 The lung bag roll cannot be 

reinstalled when the load box is in 

place. The lung bags will have to be 

set up one at a time.
5. Detach the rubber tether strap from 

the pin at the base of the neck.

Figure 4

6. Turn the body over and remove 

  the four screws that hold the two 

  halves of the upper body together. 

  These are located across the  back 

  of the shoulders. 

(See figure 4.)

Figure 5

7. Lift the top half of the body slightly 

  away from the bottom, releasing 

  the tab that holds the neck in place. 

 

(See figure 5.)

8. Remove the head and neck from 

  the body and remove the chest skin.

Figure 6

Installing the Load Box

(See figure 6.)

The next step is to install the load box 

for the defibrillation chest skin. The 

load box is the box attached to the 

skin via one large cable and a smaller 

cable with four lead snaps attached.
This is the device that will absorb the 

energy being administered to the 

chest skin during defibrillation.

Note:

 It is very important that this 

load box is installed properly!

*RESUSCI

®

 JUNIOR IS A TRADEMARK OF LAERDAL MEDICAL CORPORATION.

2

Note:

 

If the defibrillation chest skin 

is not functioning or wiring comes 

undone, please contact us to repair or 

replace the unit. Failure to do so, or 

unauthorized repair, may void the war-

ranty or cause further harm or dam-

age to you or your equipment.

Accessories

LF03656U 

Physio Control 
Defribrillation Pad and 
Patient Simulator 
Adapter Package for 
LifePak 10 and Lifepak 20

LF03657U

  Marquette Electronics 

Defribrillation Pad and 
Patient Simulator Adapter 
Package

LF03658U

  SpaceLabs/Laerdal/

Heartstart/First Medic
Defibrillation Pad 
and Patient Simulator
Adapter Package

LF03961U

  Zoll Training Cables with 

 

Adapters

LF03962U

  Physio Control Training 

 

Cables with Adapters

 

for LifePak 12

If you need help selecting the training 

pad adapters that correspond to your 

AED unit, please feel free to call us at 

1-800-558-9595 for assistance.

5

NP016_96_LF03611   3

NP016_96_LF03611   3

10/5/12   10:24:12 AM

10/5/12   10:24:12 AM

Summary of Contents for LF03611U

Page 1: ...truction Manual Products by Nasco Q ment ment ement kin a Tutor ement fusion on a Stand ement anikin kin CPR Manikin Cricoid omplete nikin Manikin ltation ltation mulator kin Series eous hild Manikin n and 10 5 12 10 24 12 AM 10 5 12 10 24 12 AM ...

Page 2: ...ttle of Nasco Cleaner Load Box Mounting Plate 2 Machine Screws Figure 2 Things You Will Need Phillips Screwdriver Note 360 joules is the maximum energy level that Nasco recommends administering to the defibrillation chest skin Energy levels in excess of 360 joules may cause irreparable damage to the chest skin circuitry and patient simulator being used thus voiding Nasco s warranty and endangering...

Page 3: ...remove the four screws that hold the two halves of the upper body together These are located across the back of the shoulders See figure 4 Figure 5 7 Lift the top half of the body slightly away from the bottom releasing the tab that holds the neck in place See figure 5 8 Remove the head and neck from the body and remove the chest skin Figure 6 Installing the Load Box See figure 6 The next step is ...

Page 4: ... cable It can also be attached to an Arrhythmia Tutor which is available from Nasco through the Health Care Educational Materials Catalog For now simply uncoil the cable and allow it to hang freely from the manikin To reinstall the head simply reverse the procedure used to remove it Make sure to tighten the four body screws Figure 10 Next replace the chest compression plate that was removed earlie...

Page 5: ... tor doesn t have the standard snap connectors Please see accessories at the end of this manual Once your manikin is connected to your patient simulator you will be able to pick up the ECG waves either through the monitor hook ups on the skin or through the two disks attached to the skin on the defibrillation sites These disks will enable you to pick up the ECG wave using either the Quick Look pad...

Page 6: ... to you or your equipment Accessories LF03656U Physio Control Defribrillation Pad and Patient Simulator Adapter Package for LifePak 10 and Lifepak 20 LF03657U Marquette Electronics Defribrillation Pad and Patient Simulator Adapter Package LF03658U SpaceLabs Laerdal Heartstart First Medic Defibrillation Pad and Patient Simulator Adapter Package LF03961U Zoll Training Cables with Adapters LF03962U P...

Page 7: ...le Female Catheter LF01155U Advanced CPR Dog LF01162U Venatech IV Trainer LF01174U NG Tube Trach Skills LF01184U Venatech IM Sub Q LF01193U Special Needs Baby LF03000U CPARLENE Series LF03601U Adult Airway Management Trainer with Stand LF03602U Adult Airway Management Manikin LF03609U Child Airway Management Trainer with Stand LF03616U Child CRiSis Manikin LF03617U Deluxe Child CRiSis Manikin with...

Reviews:

Related manuals for LF03611U