background image

Appendix B 

Troubleshooting 

Observation 

Possible cause 

Recommended action 

The LCD screen remains 
blank and the fan does not 
run when the power is 

turned on 

AC power cord is not connected 

Check AC power cord connections at both ends.  
Use the correct cord. 

The fuse has blown 

Replace the fuse 

Operation stops with 
alarm: The screen displays 
“NO LOAD” 

Electrophoresis leads are not 
connected to the power supply or 
to the electrophoresis unit(s), or 

there is a broken circuit in the 
electrophoresis cell 

Check the connections to the power supply and 
on your electrophoresis cell to make sure the 
connection is intact; check condition of wires in 

electrophoresis unit. Close the circuit by 
reconnecting the cables. Press Start/Pause to 
restart the run. 

High resistance due to tape left on 
a pre-cast gel, incorrect buffer 

concentration, or incorrect buffer 
volumes in the electrophoresis 
cell 

Correct the condition by making sure the tape 
is removed from the pre-cast gel, buffers are 

prepared correctly, and the recommended 
volume of buffer is added to the electrophoresis 
unit. 

Operation stops with 
alarm: Display shows 
“OVER VOLTAGE” 

Circuit is interrupted 

 

Verify that the running buffer is correct. 

 

Verify the all cables are attached correctly 

 

Turn the Power switch off and on again; 
restart application. 

 

If you cannot restart the instrument, turn 
off the power, disconnect the power cord 
from the outlet, and contact Technical 
Service. 

Operation stops with 
alarm: Display 
shows“LEAKAGE” 

Ground leak detected during run 

Check the electrophoresis system for improper 
grounding. Restart the power supply by turning 
the Power switch off and on. 

Operation stops with 
alarm: 
Display shows “OVER 
TEMP” 

Power supply is overheating 

 

Turn off power supply. Check for sufficient 
airflow around the power supply fan. After 
cooling down, restart the power supply by 
turning the Power switch to the on position. 

 

If you cannot restart the instrument, turn 
off the power, disconnect the power cord 
from the outlet, and contact Technical 
Service. 

 

 

 

PowerEase

®

 90W Power Supply User Guide

  

 

13

 

 

Summary of Contents for PowerEase 90W

Page 1: ...user guide Research Use Only Not for use in diagnostic procedures PowerEase 90W Power Supply Catalog Numbers PS0090 PS0091 PS0092 Publication Number MAN0009597 Revision B 0 ...

Page 2: ...ANTABILITY FITNESS FOR A PARTICULAR PURPOSE OR NON INFRINGEMENT TO THE EXTENT ALLOWED BY LAW IN NO EVENT SHALL LIFE TECHNOLOGIES AND OR ITS AFFILIATE S BE LIABLE WHETHER IN CONTRACT TORT WARRANTY OR UNDER ANY STATUTE OR ON ANY OTHER BASIS FOR SPECIAL INCIDENTAL INDIRECT PUNITIVE MULTIPLE OR CONSEQUENTIAL DAMAGES IN CONNECTION WITH OR ARISING FROM THIS DOCUMENT INCLUDING BUT NOT LIMITED TO THE USE ...

Page 3: ... parameter settings 8 Appendix A 10 General safety 10 Instrument safety 10 Explanation of symbols and warnings 11 Production regulatory compliance 12 Appendix B 13 Troubleshooting 13 Common errors 14 Appendix C 15 Operational electrical parameters 15 Appendix D 16 Product specifications 16 Appendix E 17 Repair and maintenance 17 Appendix F 18 Documentation and support 18 PowerEase 90W Power Supply...

Page 4: ...rrier keep the supplied box for inspection The warranty does not cover in transit damage To ensure safe reliable operation always operate the PowerEase 90W Power Supply in accordance with the manufacturer s instructions Always wear protective gloves and safety glasses when working in a laboratory environment Avoid exposing the power supply to liquids as these may damage the unit and void the warra...

Page 5: ...upply Constant Voltage or Current Capable of running multiple electrophoresis units LCD display with clear menu prompts for easy use Four sets of output terminals Capability to specify run durations PowerEase 90W Power Supply User Guide 5 ...

Page 6: ...ursor position down and decrease values when settings are flashing Mode Cycle between voltage current and time settings allowing values to be changed Constant Toggle between constant voltage and current operation Start Pause Start pause and resume power supply operation without resetting the timer Stop Stop power supply operation reset the timer turn off the beeper Control buttons Output jacks Bac...

Page 7: ...ed in this section We recommend that you carefully review these guidelines before operating the instrument Important For best results do NOT use the PowerEase 90W Power Supply at its maximum electrical load limits Variations in buffer conditions can result in exceeding the power supply s maximum voltage or current output capacity and produce undesirable variations in performance Recommendation The...

Page 8: ...rge bold type on the left side of the display and TIME and AMP in small type on the right side of the display 2 Press the Constant key to select either Constant Voltage or Constant Current from the display screen The screen displays the selected operating mode in large bold type on the left of the display 3 Use the arrow keys to set the selected operation voltage VOLT or current AMP to the appropr...

Page 9: ...sistance during the run If you use the current limiting setting on the PowerEase 90W Power Supply the voltage will increase as resistance increases to satisfy Ohm s law V IR If no voltage limit is set and a local fault condition occurs such as a poor connection very high local resistance may cause the voltage to increase to the maximum capacity of the power supply This may lead to local overheatin...

Page 10: ...en RUNNING is displayed on the screen 2 WAIT at least 5 seconds after stopping a run before handling output leads or connected apparatus 3 ALWAYS make sure that hands work area and instruments are clean and dry before making any connections or operating the power supply 4 ALWAYS connect the power supply to a properly grounded AC outlet Avoid damage to the instrument 1 For proper ventilation leave ...

Page 11: ...es négatives sur l environnement à la suite de l élimination de déchets électroniques ne pas éliminer ce déchet électronique avec les déchets usuels non soumis au tri sélectif Se conformer aux ordonnances locales sur les déchets municipaux pour les dispositions d élimination et communiquer avec le service à la clientèle pour des renseignements sur les options d élimination responsable Indicates co...

Page 12: ...sed in accordance with the instruction manual may cause harmful interference to radio communications Operation of this equipment in a residential area is likely to cause harmful interference in which case the user will be required to correct the interference at their expense Changes or modifications not expressly approved by the party responsible for compliance could void the user s authority to o...

Page 13: ...he tape is removed from the pre cast gel buffers are prepared correctly and the recommended volume of buffer is added to the electrophoresis unit Operation stops with alarm Display shows OVER VOLTAGE Circuit is interrupted Verify that the running buffer is correct Verify the all cables are attached correctly Turn the Power switch off and on again restart application If you cannot restart the instr...

Page 14: ...ected system Excessive temperature increase Excessive buffer evaporation Loose connection in a connected system Current set too low Change in constant mode Voltage changes to current Current set too low Ceiling hit and constant mode changed from voltage to current Increase current to 500 mA Current changes to voltage Voltage set to low Ceiling hit and constant mode changed from current to voltage ...

Page 15: ...re gradually replaced by the slower moving less conductive glycine ions from the running buffer as the gel runs As a result the resistance of the gel increases as the chloride glycine front moves down the gel Voltage The velocity with which an ion moves in an electric field will vary in proportion to the field strength volts per unit distance The higher the voltage the faster an ion will move Curr...

Page 16: ...emental settings 1 mA incremental settings Crossover Automatic Display type Backlit LCD Graphic type Display size 53 64 x 15 64 mm W x H Pause function Yes Safety features No Load Detection Load Change Detection Overload Detection Ground Leak Detection Auto Restart Stackable Yes Housing material Flame retardant ABS Housing size 200 x 290 x 70 mm W x D x H Operating temperature 0 C 40 C Environment...

Page 17: ...he fuse 1 Turn off the main power switch at the rear of the PowerEase 90W Power Supply and detach the power cord from the rear of the PowerEase 90W Power Supply 2 Open the fuse compartment located inside the Power Inlet by inserting a small flat blade screwdriver into the slot on top of the fuse compartment Turn the screwdriver to gently pry open the fuse compartment Note The fuse compartment will...

Page 18: ... and patches Safety Data Sheets SDS Safety Data Sheets SDSs are available at www lifetechnologies com support Certificate of Analysis The Certificate of Analysis provides detailed quality control and product qualification information for each product Certificates of Analysis are available on our website Go to www lifetechnologies com support and search for the Certificate of Analysis by product lo...

Page 19: ......

Page 20: ...For support visit lifetechnologies com support or email techsupport lifetech com lifetechnologies com 06 May 2014 ...

Reviews: