Operation modes 

Introduction 

The PowerEase

®

 90W Power Supply is a microprocessor-controlled power supply 

designed to meet most electrophoresis needs in a single, easy to use unit. The power 
supply offers two modes: Constant Voltage and Constant Current. Use the Constant 
Voltage or Constant Current Mode for applications that require only one specific voltage 
limit or current limit during the entire duration of electrophoresis. 

 

Constant operation 
protocol 

Instructions for operating the PowerEase

®

 90W Power Supply in the 

Constant Mode 

are 

provided in this section.  The 

Constant Voltage 

or

 Constant Current Mode 

allows you to 

specify a voltage limit or current limit to be used during the entire duration of 
electrophoresis.  Review the guidelines provided in this manual before starting 
electrophoresis using the PowerEase

®

 90W Power Supply. 

A basic 

Constant Voltage 

or 

Constant Current Mode 

operating procedure of the 

PowerEase

®

 90W Power Supply is provided below. We recommend reading the 

guidelines provided in this manual for best results before starting an operation. 

1.

 

Use the power switch on the rear side of the instrument to turn on the PowerEase

®

 

90W Power Supply.  

The display screen will illuminate. 

The screen displays the last set 

(voltage) in large bold type on the left side of the 

display and 

TIME

 and 

AMP

 in small type on the right side of the display. 

2.

 

Press the 

Constant

 key to select either 

Constant Voltage

 or 

Constant Current

 from 

the display screen. 

The screen displays the selected operating mode in large bold type on the left of the 
display.  

3.

 

Use the 

 

 arrow keys to set the selected operation [voltage (VOLT) or current 

(AMP)] to the appropriate values. 

4.

 

Use the 

Mode

 key to select 

TIME

 and use 

 

 arrow key to set the time 

(hours/minutes) to specify the duration of the electrophoresis.  

5.

 

Press 

Start/Pause 

key to start electrophoresis.  The Red LED illuminates when the 

electrophoresis is running. 

6.

 

Press the 

Start/Pause

 key again to temporarily interrupt power to ongoing 

electrophoresis without terminating the operation along with LED flashing. 

7.

 

Press 

Start/Pause 

key to restart electrophoresis. 

8.

 

Press 

Stop 

key to stop the electrophoresis. 

Note:

 If the parameters need to be changed during the run you must press 

Start/Pause

, enter the 

changes, and then press 

Start/Pause

 once again to restart your operation. 

Note

: After stopping and restarting an operation, the timer resets to the selected time  

 

Limiting parameter settings 

Introduction 

The PowerEase

®

 90W Power Supply is capable of operating at limiting Voltage, or 

limiting Current.   We recommend operating the PowerEase

®

 90W Power Supply at 

limiting voltage for most applications. See below for more details. 

 

 

PowerEase

®

 90W Power Supply User Guide

 

Summary of Contents for PowerEase 90W

Page 1: ...user guide Research Use Only Not for use in diagnostic procedures PowerEase 90W Power Supply Catalog Numbers PS0090 PS0091 PS0092 Publication Number MAN0009597 Revision B 0 ...

Page 2: ...ANTABILITY FITNESS FOR A PARTICULAR PURPOSE OR NON INFRINGEMENT TO THE EXTENT ALLOWED BY LAW IN NO EVENT SHALL LIFE TECHNOLOGIES AND OR ITS AFFILIATE S BE LIABLE WHETHER IN CONTRACT TORT WARRANTY OR UNDER ANY STATUTE OR ON ANY OTHER BASIS FOR SPECIAL INCIDENTAL INDIRECT PUNITIVE MULTIPLE OR CONSEQUENTIAL DAMAGES IN CONNECTION WITH OR ARISING FROM THIS DOCUMENT INCLUDING BUT NOT LIMITED TO THE USE ...

Page 3: ... parameter settings 8 Appendix A 10 General safety 10 Instrument safety 10 Explanation of symbols and warnings 11 Production regulatory compliance 12 Appendix B 13 Troubleshooting 13 Common errors 14 Appendix C 15 Operational electrical parameters 15 Appendix D 16 Product specifications 16 Appendix E 17 Repair and maintenance 17 Appendix F 18 Documentation and support 18 PowerEase 90W Power Supply...

Page 4: ...rrier keep the supplied box for inspection The warranty does not cover in transit damage To ensure safe reliable operation always operate the PowerEase 90W Power Supply in accordance with the manufacturer s instructions Always wear protective gloves and safety glasses when working in a laboratory environment Avoid exposing the power supply to liquids as these may damage the unit and void the warra...

Page 5: ...upply Constant Voltage or Current Capable of running multiple electrophoresis units LCD display with clear menu prompts for easy use Four sets of output terminals Capability to specify run durations PowerEase 90W Power Supply User Guide 5 ...

Page 6: ...ursor position down and decrease values when settings are flashing Mode Cycle between voltage current and time settings allowing values to be changed Constant Toggle between constant voltage and current operation Start Pause Start pause and resume power supply operation without resetting the timer Stop Stop power supply operation reset the timer turn off the beeper Control buttons Output jacks Bac...

Page 7: ...ed in this section We recommend that you carefully review these guidelines before operating the instrument Important For best results do NOT use the PowerEase 90W Power Supply at its maximum electrical load limits Variations in buffer conditions can result in exceeding the power supply s maximum voltage or current output capacity and produce undesirable variations in performance Recommendation The...

Page 8: ...rge bold type on the left side of the display and TIME and AMP in small type on the right side of the display 2 Press the Constant key to select either Constant Voltage or Constant Current from the display screen The screen displays the selected operating mode in large bold type on the left of the display 3 Use the arrow keys to set the selected operation voltage VOLT or current AMP to the appropr...

Page 9: ...sistance during the run If you use the current limiting setting on the PowerEase 90W Power Supply the voltage will increase as resistance increases to satisfy Ohm s law V IR If no voltage limit is set and a local fault condition occurs such as a poor connection very high local resistance may cause the voltage to increase to the maximum capacity of the power supply This may lead to local overheatin...

Page 10: ...en RUNNING is displayed on the screen 2 WAIT at least 5 seconds after stopping a run before handling output leads or connected apparatus 3 ALWAYS make sure that hands work area and instruments are clean and dry before making any connections or operating the power supply 4 ALWAYS connect the power supply to a properly grounded AC outlet Avoid damage to the instrument 1 For proper ventilation leave ...

Page 11: ...es négatives sur l environnement à la suite de l élimination de déchets électroniques ne pas éliminer ce déchet électronique avec les déchets usuels non soumis au tri sélectif Se conformer aux ordonnances locales sur les déchets municipaux pour les dispositions d élimination et communiquer avec le service à la clientèle pour des renseignements sur les options d élimination responsable Indicates co...

Page 12: ...sed in accordance with the instruction manual may cause harmful interference to radio communications Operation of this equipment in a residential area is likely to cause harmful interference in which case the user will be required to correct the interference at their expense Changes or modifications not expressly approved by the party responsible for compliance could void the user s authority to o...

Page 13: ...he tape is removed from the pre cast gel buffers are prepared correctly and the recommended volume of buffer is added to the electrophoresis unit Operation stops with alarm Display shows OVER VOLTAGE Circuit is interrupted Verify that the running buffer is correct Verify the all cables are attached correctly Turn the Power switch off and on again restart application If you cannot restart the instr...

Page 14: ...ected system Excessive temperature increase Excessive buffer evaporation Loose connection in a connected system Current set too low Change in constant mode Voltage changes to current Current set too low Ceiling hit and constant mode changed from voltage to current Increase current to 500 mA Current changes to voltage Voltage set to low Ceiling hit and constant mode changed from current to voltage ...

Page 15: ...re gradually replaced by the slower moving less conductive glycine ions from the running buffer as the gel runs As a result the resistance of the gel increases as the chloride glycine front moves down the gel Voltage The velocity with which an ion moves in an electric field will vary in proportion to the field strength volts per unit distance The higher the voltage the faster an ion will move Curr...

Page 16: ...emental settings 1 mA incremental settings Crossover Automatic Display type Backlit LCD Graphic type Display size 53 64 x 15 64 mm W x H Pause function Yes Safety features No Load Detection Load Change Detection Overload Detection Ground Leak Detection Auto Restart Stackable Yes Housing material Flame retardant ABS Housing size 200 x 290 x 70 mm W x D x H Operating temperature 0 C 40 C Environment...

Page 17: ...he fuse 1 Turn off the main power switch at the rear of the PowerEase 90W Power Supply and detach the power cord from the rear of the PowerEase 90W Power Supply 2 Open the fuse compartment located inside the Power Inlet by inserting a small flat blade screwdriver into the slot on top of the fuse compartment Turn the screwdriver to gently pry open the fuse compartment Note The fuse compartment will...

Page 18: ... and patches Safety Data Sheets SDS Safety Data Sheets SDSs are available at www lifetechnologies com support Certificate of Analysis The Certificate of Analysis provides detailed quality control and product qualification information for each product Certificates of Analysis are available on our website Go to www lifetechnologies com support and search for the Certificate of Analysis by product lo...

Page 19: ......

Page 20: ...For support visit lifetechnologies com support or email techsupport lifetech com lifetechnologies com 06 May 2014 ...