2
2
31
Ventilations (Green Light):
Adult Setting:
The ventilation light
will turn on every time the ventilation
of 0.80 liters is reached. It will turn off
when volume drops below this level.
Child Setting:
The ventilation light
will turn on every time the ventilation
of 0.50 liters is reached. It will turn off
when volume drops below this level.
Figure 3
Figure 4
2. Check alignment of lower slide
and upper groove reader inside
chest cavity.
3. Turn power on and test.
The ventilation volume indicator
reports low or no levels.
1. Turn power off.
2. Check to see if the lung bag is
properly installed. Lung bag may
be blocking the ventilation reader.
3. Check to see that ventilation
reader slides easily.
4. Turn power on.
Compressions (Yellow Light):
Adult Setting:
The compression light
will turn on every time the compres-
sion depth of 1.50 inches is reached.
It will turn off when it drops below this
depth or if it exceeds 2.00 inches.
Child Setting:
The compression light
will turn on every time the compres-
sion depth of 1.00 inch is reached. It
will turn off when it drops below this
depth or if it exceeds 1.50 inches
TROUBLE SHOOTING
The red placement light is continu-
ously on indicating improper hand
placement.
1. Turn power off.
2 The chest plate sensor has
become disconnected from the
red connector in the chest cavity.
(See figure 3.)
3 Check inside chest cavity to see if
electronic bar reader assembly is
in proper vertical alignment.
(See figure 4.)
4. Turn power on and test.
The compression depth indicator
reports only low compressions.
1. Turn power off.
PROCEDURE FOR MEMORY UNIT:
Setup Battery Installation
Remove the cover using the key
provided and inserting in the four
open slots on the sides of the unit
(two on each side).
(See figure 5.)
Press inward while gently applying
upward pressure until you hear a soft
click. If the supplied key is not avail-
able, a small screwdriver can be used.
Batteries are included. All electronic
components use “D” cell alkaline
batteries. On models requiring six
batteries, you may find it easier to
remove or install batteries if the center
two batteries are removed first and
installed last.
Procedure D
Procedure E