48
DM00003 Rev D
SYSTEM
MANUAL
Recommended separation distances between
portable and mobile RF communications equipment and the Trackit M
The Trackit M is intended for use in an electromagnetic environment in which radiated FR disturbances
are controlled. The customer or the user of the Trackit M can help prevent electromagnetic interference by
maintaining a minimum distance between portable and mobile RF communications equipment (transmitters)
and the Trackit M as recommended below, according to the maximum output power of the communications
equipment.
Rated maximum output
power
of transmitter
W
Separation distance according to frequency of transmitter
m
150 kHz to 80 MHz
d = 1,2√
P
80 MHz to 800 MHz
d = 1,2√
P
800 MHz to 2.5 GHz
d = 2,3√
P
0,01
0,12
0,12
0,23
0,1
0,38
0,38
0,73
1
1,2
1,2
2,3
10
3,8
3,8
7,3
100
12
12
23
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in
metres (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the
maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propogation is affected by absorption
and reflection from structures, objects and people.