DISPLAY MESSAGES AND PROBLEM-SOLVING GUIDE
85
Here are answers to some
questions you might ask about
your InDuo™ insulin doser display:
Why has the display turned off?
If the display goes off, press the
release button, and the push button
will jump out. Press the push button
completely in, and the display
should reappear. If the display does
not reappear, open and close the
slide. Prime the InDuo™ insulin
doser as described on page 24. If
the display remains off, the insulin
doser is not working. Do not use the
insulin doser, instead call the InDuo™
Customer Care Line at 1 888 729-2299
to replace your insulin doser.
Why is the display showing a minus
symbol?
The dose selector has been dialed
below zero
0
. Dial forward until the
display shows the correct number of
units.
Why does the display not respond
when I turn the dose selector?
Press the release button so that the
push button jumps out before a dose
can be dialed.
On the next page are the answers
to some questions you might ask
about priming the insulin doser
and injecting insulin:
DCO_06153701A_ID_OB_CAN 9/6/02 11:37 AM Page 85