background image

DISPLAY SYMBOLS AND MESSAGES

Error message could be caused by a used
test strip or indicates a temporary or
permanent electronics problem*

Error message. The blood or control
solution sample was applied before 

R

appeared on the display

*

Three possible reasons:

1. Possible high glucose in a test

environment near low end of operating
temperature range (6°–44°C/43°–111°F).
Retest in a warmer environment. Or;

2. Test strip damaged or moved during

testing. Repeat test with new test strip.
Or;

3. Sample improperly applied. Review

blood application and repeat test with
new test strip.

Test strip damaged or confirmation
window not full

*

* Retest with a new test strip. If the problem

persists, call the InDuo™ Customer Care Line 
at 1 888 729-2299.

Summary of Contents for novo nordisk InDuo

Page 1: ...Quick Reference Guide See inside for instructions on using InDuo Important Read detailed instructions in your InDuo user manual...

Page 2: ...h Ultra Test Strips OneTouch Ultra Control Solution OneTouch UltraSoft Adjustable Blood Sampler OneTouch UltraSoft Sterile Lancet OneTouch UltraClear Cap for alternate puncture site Penfill 3 ml insul...

Page 3: ...eter display with code number on test strip vial If matching begin testing If not matching see Step 3 Step 1 Enter code mode Press until codes match After the apply symbol R appears meter is ready for...

Page 4: ...meter turns on automatically Touch and hold blood drop to top edge of test strip where it meets narrow channel Hold blood drop to top edge until confirmation window is completely filled before meter...

Page 5: ...wer than 1 1 mmol L 20 mg dL Treat this condition according to healthcare professional recommendation The temperature of the air blood glucose meter or test strips was below the operating range of the...

Page 6: ...averages Battery power is too low to run a test Replace the battery immediately The time flashing on the display indicates the meter lost power briefly and entered setting mode Reset time time format...

Page 7: ...ssible high glucose in a test environment near low end of operating temperature range 6 44 C 43 111 F Retest in a warmer environment Or 2 Test strip damaged or moved during testing Repeat test with ne...

Page 8: ...Fine 6 or 8 mm needle To prime the InDuo insulin doser press the release button the push button will jump out Dial 1 unit For new Penfill dial 8 units Hold the insulin doser with the needle pointing u...

Page 9: ...e push button completely in until it locks After the injection discard the needle carefully Always replace the insulin doser into the meter cap Press the release button and the push button jumps out D...

Page 10: ...Duo insulin doser The display shows your previous dose and the time passed since delivery Example 1 You injected 6 units 2 hours ago Example 2 You injected 70 units 10 hours ago Example 3 You injected...

Page 11: ...is intact e g no cracks If the insulin is a suspension cloudy insulin re suspend the insulin as described in the Penfill package insert Insert the new Penfill into the InDuo insulin doser by pushing i...

Page 12: ...For assistance call the InDuo Customer Care Line at 1 888 729 2299 Manufactured by LifeScan Inc Milpitas California 95035 U S A LifeScan Novo Nordisk 2002 Revision date 06 02...