background image

22

CONTROL SOLUTION TESTING

  3  Control solution testing

Control solution testing precautions

OneTouch Select

®

 Plus Control Solution is used to check that 

the meter and test strips are working together properly and that 
the test is performing correctly. (Control solution is available 
separately.)

NOTE: 

•  When you first open a new vial of control solution, record  

the discard date on the vial label. Refer to the control  
solution insert or vial label for instructions on determining 
the discard date. 

•  Tightly close the cap on the control solution vial immediately 

after use to avoid contamination or damage.

• 

Do Not

 open the test strip vial until you are ready to remove 

a test strip and perform a test. Use the test strip 

immediately

 

after removing it from the vial, especially in high humidity 
environments. 

•  Control solution tests must be done at room temperature  

(20-25°C). Make sure your meter, test strips and control 
solutions are at room temperature before testing. 

CAUTION: 

• 

Do Not

 swallow or ingest control solution.

• 

Do Not

 apply control solution to the skin or eyes as it may 

cause irritation.

• 

Do Not

 use control solution after the expiry date (printed on 

the vial label) or the discard date, whichever comes first, or 
your results may be inaccurate. 

Do a control solution test

•  If you suspect that the meter or test strips are not working 

properly.

•  If you have had repeated unexpected blood glucose results.
•  If you drop or damage the meter.

Summary of Contents for OneTouch Select Plus Simple

Page 1: ...Blood Glucose Monitoring System Owner s Guide Simply insert a test strip to turn the meter on and begin testing Instructions for Use...

Page 2: ...C Blood Glucose Meter Test Strips and Control Solution MDD 93 42 EEC Kindly contact Customer Care at 1800 225544 or visit www OneTouch in 2018 2020 LifeScan IP Holdings LLC Meter Made in China Rev Dat...

Page 3: ...ing to know your system 1 2 Blood glucose testing 6 3 Control solution testing 22 4 Caring for your system 27 5 Battery 29 6 Troubleshooting 31 7 Detailed information about your system 37 Table of con...

Page 4: ...Warnings Refer to the Owner s Guide and inserts that came with your system for safety related information Direct current Consult Instructions for Use Manufacturer Lot Number Serial Number Storage Temp...

Page 5: ...e with the OneTouch Select Plus Test Strips OneTouch Select Plus Control Solution and the OneTouch Delica Plus Lancing Device Intended use The OneTouch Select Plus Simple Blood Glucose Monitoring Syst...

Page 6: ...n critically ill patients patients in shock dehydrated patients or hyperosmolar patients Test principle Glucose in the blood sample mixes with the enzyme Glucose Oxidase in the test strip and a small...

Page 7: ...SYSTEM 1 Getting to know your system Your OneTouch Select Plus Simple Blood Glucose Monitoring System Included with your kit OneTouch Select Plus Simple Meter CR2032 lithium coin cell battery included...

Page 8: ...quired but may not be included in your kit Use only OneTouch Select Plus Control Solutions and Test Strips with the OneTouch Select Plus Simple Meter Lancets Lancets OneTouch Select Plus Control Solut...

Page 9: ...neTouch Select Plus Simple Blood Glucose Monitoring System Meter 1 Test strip port 2 Battery icon 3 Last result icon 4 mg dL is the pre set unit of measure 5 Colour Range Indicator Bars 6 Range Indica...

Page 10: ...4 GETTING TO KNOW YOUR SYSTEM Getting to know your OneTouch Select Plus Test Strip Test strip 1 Edge to apply sample 2 Confirmation window 3 Contact bars Insert into test strip port...

Page 11: ...a Range Indicator Arrow pointing to a corresponding Colour Range Indicator Bar below the meter display Use the Range Indicator Arrow and Colour Bar to interpret your results A Range Indicator Arrow wi...

Page 12: ...emove a test strip and perform a test Use the test strip immediately after removing it from the vial especially in high humidity environments Tightly close the cap on the vial immediately after use to...

Page 13: ...n the patient s whole blood sample as it may cause inaccurate results A haematocrit percentage of your blood that is red blood cells that is either very high above 55 or very low below 30 can cause fa...

Page 14: ...8 BLOOD GLUCOSE TESTING Getting to know your OneTouch Delica Plus Lancing Device 1 Release button 2 Slider control 3 Depth indicator 4 Depth wheel 5 Lancing device cap 6 Protective cover...

Page 15: ...orm Alternate Site Testing AST The OneTouch Delica Plus Lancing System should not be used on the forearm or palm with the OneTouch Select Plus Simple Blood Glucose Monitoring System Lancing precaution...

Page 16: ...rotective cover one full turn until it separates from the lancet Save the protective cover for lancet removal and disposal See Removing the used lancet 2 Insert a sterile lancet into the lancing devic...

Page 17: ...mbers 1 to 7 on the Depth Wheel indicates an additional available depth setting Adjust the depth by turning the depth wheel Smaller numbers are for a shallower puncture and larger numbers for a deeper...

Page 18: ...is working properly Your last result will then appear and the last result icon will blink If this is your first time using the meter dashes will appear instead of a result See Error and other message...

Page 19: ...and dry completely Contaminants on the skin may affect results 1 Puncture your finger Hold the lancing device firmly against the side of your finger Press the release button Remove the lancing device...

Page 20: ...BLOOD GLUCOSE TESTING Applying the sample Line up the test strip with the blood drop Channel NOTE Do Not apply blood on the top of the test strip Gently touch the channel to the edge of the blood dro...

Page 21: ...new test strip Do Not smear or scrape the drop of blood with the test strip Do Not apply more blood to the test strip after you have moved the drop of blood away Do Not move the test strip in the met...

Page 22: ...to indicate if your result is below above or within the meter s range limits see The Range Indicator feature in Section 1 The arrow will point to the appropriate Colour Range Indicator Bar on the met...

Page 23: ...irm that the unit of measure mg dL is displayed If your display shows mmol L rather than mg dL stop using the meter and contact Customer Care NOTE If you get an error message instead of a result see L...

Page 24: ...N Dehydration and low results You may get false low results if you are severely dehydrated If you think you are severely dehydrated contact your healthcare professional immediately HI is displayed whe...

Page 25: ...h your results and you have followed all instructions in this Owner s Guide call your healthcare professional Never ignore symptoms or make significant changes to your diabetes management programme wi...

Page 26: ...ice has an ejection feature so you do not have to pull out the used lancet 1 Remove the lancing device cap Remove the cap by rotating it and then pulling it straight off the device 3 Eject the lancet...

Page 27: ...oid unintended lancet stick injuries Used lancets and test strips may be considered biohazardous waste in your area Be sure to follow your healthcare professional s recommendations or local regulation...

Page 28: ...he test strip vial until you are ready to remove a test strip and perform a test Use the test strip immediately after removing it from the vial especially in high humidity environments Control solutio...

Page 29: ...test strip port and contact bars facing you Contact bars Test strip port 2 Wait for the flashing blood drop icon to appear on the display Preparing the control solution 1 Before removing the cap shake...

Page 30: ...the small well on the top of the cap or onto another clean non absorbent surface Applying the control solution 1 Hold the meter so that the channel at the top edge of the test strip is at a slight an...

Page 31: ...along with the History Mode icon Checking if the result is in range Each vial of test strips has the OneTouch Select Plus Mid Control Solution range printed on its label Compare the result displayed o...

Page 32: ...est strips and or control solution were not all at the same temperature when the control solution test was performed A problem with the meter Dirt or contamination in the small well on the top of the...

Page 33: ...reduce your exposure to disease Cleaning your meter lancing device and cap The meter lancing device and cap should be cleaned whenever they are visibly dirty and before disinfection Clean your meter...

Page 34: ...containing a minimum of 5 5 sodium hypochlorite as the active ingredient Prepare a solution of 1 part household bleach and 9 parts water Follow manufacturer s instruction for handling and storage of...

Page 35: ...ST BATTERY This product contains a coin button cell battery If swallowed it can quickly cause severe internal burns and can lead to death Keep new and used batteries away from children If you think ba...

Page 36: ...eter battery check that the battery is correctly installed If the meter still does not power on contact Customer Care 4 Dispose of battery Dispose of the battery according to your local environmental...

Page 37: ...hough this message could be due to a test error it is safer to treat first and then do another test Always treat according to your healthcare professional s recommendations What it means You may have...

Page 38: ...ing range 10 44 C If you do not get another HI t message you can proceed with testing What it means The meter is too cold below 10 C to perform a test What to do Move the meter and test strips to a wa...

Page 39: ...with the meter What to do Do Not use the meter Contact Customer Care What it means Error message could be caused either by a used test strip or a problem with the meter What to do Repeat the test wit...

Page 40: ...a new test strip If the error message appears again contact Customer Care What it means The sample was applied before the meter was ready What to do Repeat the test with a new test strip Apply a bloo...

Page 41: ...to do Repeat the test with a new test strip If the error message appears again contact Customer Care What it means Meter battery power is low but there is still enough battery power to perform a test...

Page 42: ...ly What it means No result in memory such as the first time use of the meter or Your meter was unable to recall the last result What to do Contact Customer Care to report this occurrence unless this i...

Page 43: ...using blood drawn from a vein 1 Therefore when comparing a meter to a lab it is important to fast for at least 8 hours It is OK to drink water Meter results can be affected by factors that do not affe...

Page 44: ...your meter results to those taken from another meter Comparing your blood glucose test results taken with this meter to your results taken from a different meter is not recommended Results may differ...

Page 45: ...um coin cell battery Biological source Aspergillus Niger Calibration Plasma equivalent Memory Last glucose test result only Operating ranges Temperature 10 44 C Relative humidity non condensing 10 90...

Page 46: ...ntrations 100 mg dL Percent and number of meter results that match the laboratory test Within 5 mg dL Within 10 mg dL Within 15 mg dL 61 3 103 168 94 6 159 168 99 4 167 168 System Accuracy Results for...

Page 47: ...the medical laboratory values at glucose concentrations at or above 100 mg dL 95 0 of the total number of samples were within 15 mg dL or 15 of the medical laboratory values Regression Statistics Samp...

Page 48: ...83 2 24 560 562 88 11 61 2 06 Results show that the greatest variability observed between test strips when tested with blood is 2 11 mg dL SD or less at glucose levels less than 100 mg dL or 2 61 CV...