42
DETAILED INFORMATION ABOUT YOUR SYSTEM
Precision
Within Run Precision (300 Venous Blood Samples Tested per
Glucose Level)
Data generated using the OneTouch Select Plus Simple
®
Meter.
Target
Glucose
(mg/dL)
Mean Glucose
(mg/dL)
Standard
Deviation
(mg/dL)
Coefficient of
Variation (%)
25
25.43
1.50
5.88
40
40.33
1.56
3.86
65
63.01
2.11
3.35
120
117.43
3.07
2.61
200
196.55
4.42
2.25
350
349.25
7.83
2.24
560
562.88
11.61
2.06
Results show that the greatest variability observed between test
strips when tested with blood is 2.11 mg/dL SD or less at glucose
levels less than 100 mg/dL, or 2.61% CV or less at glucose levels
at 100 mg/dL or above.
Total Precision (600 Control Solution Tests per
Glucose Level)
Data generated using the OneTouch Select Plus Simple
®
Meter.
Glucose Level
Ranges
(mg/dL)
Mean Glucose
(mg/dL)
Standard
Deviation
(mg/dL)
Coefficient of
Variation (%)
20 - 30
25.74
1.13
4.41
30 - 50
46.19
1.29
2.80
96 - 144
111.93
2.27
2.03
280 - 420
362.94
6.09
1.68
420 - 600
546.34
10.27
1.88