 IMPORTANT:

 

Please read this information and your OneTouch

®

 Ultra

®

 Family of Meters and 

the OneTouch

®

 Ping

 System User Guide before using blue OneTouch

®

 Ultra

®

 Test Strips. 

Do Not use your blue OneTouch

®

 Ultra

®

 Brand Test Strips if your vial is open or damaged in 

any way as this could lead to error messages or inaccurate blood glucose values. Call the 

OneTouch

®

 Customer Care Line at 1 800 663-5521 immediately if the test strip vial is open 

or damaged, or if these instructional materials or your meter results seem unclear. If you 

cannot reach Customer Service, contact your healthcare professional for advice.

In vitro diagnostic. For self-testing.

Intended Use

Blue OneTouch

® 

Ultra

®

 Test Strips are used with the OneTouch

®

 Ultra

®

 Family of Meters and 

OneTouch

®

 Ping

 System for quantitatively measuring glucose in fresh capillary whole blood. 

The blue OneTouch

®

 Ultra

®

 Test Strips and associated meters are intended for use by people 

with diabetes at home and healthcare professionals in the clinical setting. Blue OneTouch

®

 

Ultra

®

 Test Strips and associated meters are for use in fingertip, forearm, and palm testing.

 

Storage and Handling

•   Store the test strip vial in a cool dry place below 30°C. 

Do Not

 refrigerate. Keep away from 

direct sunlight and heat. Exposure to temperatures and/or humidity outside the required 
storage conditions may result in inaccurate readings. 

•   Store your test strips in their 

original vial only

. To avoid damage or contamination, 

Do 

Not

 transfer test strips to any other container.

•   

Do Not

 open the test strip vial until you are ready to test. 

Only open vial when removing 

test strips.

•   After removing a test strip from the vial, immediately close the vial lid tightly. Use each 

test strip immediately after 

removing it from the vial. 

•   

Do Not

 use test strips from any vial that is damaged or left open to air.

•   Write the discard date (date opened plus 6 months) on the vial label when you first open it.
•  

 Do  Not

 use test strips beyond the expiration (printed on vial label) or discard date, 

whichever comes first.

•   Avoid getting dirt, food, or liquids on the test strip. With clean, dry hands, you may touch 

the test strip anywhere on its surface.

•   

Do Not

 bend, cut, or alter the test strip in any way.

•   Test strips are for single use only. 

Never reuse a test strip that had blood or control 

solution applied to it.

•   Make sure your meter and test strips are about the same temperature before you test.
•   Apply only control solution or a blood sample to the test strip.
•   After performing a test, 

Do Not

 return the used test strip to the vial.

•   Used test strips may be considered biohazardous waste in your area.  Be sure to follow your 

healthcare professional’s recommendations or your local regulations for proper disposal. 

WARNING:

 

Keep the test strip vial away from children; test strips are a choking 

hazard.  Do Not swallow test strips. The test strip vial may contain drying agents that 

are harmful if inhaled or swallowed and may cause skin or eye irritation. Do Not ingest 

or swallow any items.

Blood Glucose Test Procedure

For instructions on performing a blood test (including blood sample collection), refer to 
the User Guide that came with your system. 

IMPORTANT:

 Some meters in the OneTouch

®

 Ultra

®

 Family and the OneTouch

®

 

Ping

™ 

System require coding. For meters that require coding, matching the code 

on the meter to the code on the test strip vial is essential to obtain accurate 

results. Refer to the User Guide that came with your system to determine if your 

meter requires coding and get detailed instructions on coding.

Test Results

Low Glucose Values

If your test result is below 1.1 mmol/L (20 mg/dL), a warning message will appear indicating 
a low glucose level. This may indicate severe hypoglycemia (low blood glucose). 

Treat this 

condition immediately, according to your healthcare professional’s recommendations. 

Although this message could be due to a test error, it is safer to treat first, and then do another 
test. 

High Glucose Values

If your test result is above 33.3 mmol/L (600 mg/dL), a warning message will appear 
indicating a high glucose level. This may indicate severe hyperglycemia (high blood glucose). 
You should retest your glucose level. If the message appears again, call your healthcare 
professional immediately.

If You Get Unexpected Results

If your blood glucose result is below 4.0 mmol/L (72 mg/dL)

2

, indicating low blood glucose, 

or above 10.0 mmol/L (180 mg/dL), indicating high blood glucose, you should contact and 
follow your healthcare professional’s treatment advice.

1

 If you continue to get unexpected 

results, check your system with control solution. If you are experiencing symptoms that are 
not consistent with your blood glucose test results AND you have followed all instructions 
described in your User Guide, call your healthcare professional. Never ignore symptoms  
or make significant changes to your diabetes control program without speaking to your 
healthcare professional.

Range of Expected Values

Blood glucose management requires the help of a healthcare professional. Together you can 
set your own range of expected blood glucose values, arrange your testing times, and discuss 
the meaning of your blood glucose results.
Expected blood glucose levels for people without diabetes:

2

Time 

Range, mmol/L  

              Range, mg/dL

Fasting 

Less than 6.1 

              Less than 110

2 hours after meals 

Less than 7.8 

              Less than 140

Checking the System

Use   OneTouch

®

 Ultra

®

 Control Solution

A control solution test is performed to check that the meter and test strips are working together 
properly and that you are performing the test correctly. For instructions on how and when to 
check the system by performing a control solution test, refer to the User Guide that came with 
your system. 

Limitations of Procedure

Blue OneTouch

®

 Ultra

®

 Test Strips give accurate results when the following limitations  

are observed:
•  

Do Not

 use for the diagnosis of diabetes or for testing of newborns.

•  Test strips are for single use only. 

Do Not

 reuse.

•   The test strips are specific to D-glucose and do not react to other sugars, which may be 

present in blood.

•   Use only fresh capillary whole blood. 

Do Not

 use serum or plasma.

•   Hematocrit is the percentage of red blood cells in the blood. Extremes in hematocrit 

may affect test results.

3

 Hematocrit levels below 30% may cause falsely high readings 

and hematocrit levels over 55% may cause falsely low readings. If you do not know your 
hematocrit level, consult your healthcare professional.

•   Blue OneTouch

®

 Ultra

®

 Test Strips may be used at altitudes up to 3048 meters  

(10,000 feet) without an effect on test results. Accurate results were demonstrated in 
clinical studies performed at altitudes up to 1609 meters (5,280 feet) and in studies 
simulating altitudes up to 3048 meters (10,000 feet).

Healthcare professionals—please note these additional limitations of procedure:
•   Fresh capillary blood may be collected into heparin-containing test tubes if the blood is 

used within 10 minutes. 

Do Not

 use other anticoagulants or preservatives.

•   Interferences: Acetaminophen, salicylates, uric acid, ascorbic acid (vitamin C), and 

other reducing substances (when occurring in normal blood or normal therapeutic 
concentrations) do not significantly affect results. However, abnormally high concentrations 
in blood may cause inaccurately high results.

•   Patients undergoing oxygen therapy may yield falsely low results.
•   Test results may be falsely low if the patient is severely dehydrated, in shock, or in a 

hyperosmolar state (with or without ketosis). Critically ill patients should not be tested by 
blood glucose meters.

•   Lipemic samples: Cholesterol levels up to 18.1 mmol/L (700 mg/dL) and triglycerides up to 

33.9 mmol/L (3000 mg/dL) do not affect the results. Grossly lipemic patient samples have 
not been tested and are not recommended for testing with the OneTouch

®

 Ultra

®

 Family of 

Meters and the OneTouch

®

 Ping

 System.

 

Test Principle

The OneTouch

®

 Ultra

®

 Family of Meters and OneTouch

®

 Ping

 System are plasma-calibrated to 

allow easy comparison of results with laboratory methods. Glucose in the blood sample mixes 
with special chemicals on the test strip and a small electrical current is produced. This current 
is measured by the OneTouch

®

 Ultra

®

 Family of Meters and OneTouch

®

 Ping

 System and 

displayed as your blood glucose result. The strength of this current changes with the amount 
of glucose in the blood sample.

Reagent Composition

Each test strip contains: Glucose oxidase (Aspergillus niger) 

 0.08 IU; ferricyanide

 

 22 µg; 

other ingredients (buffer, etc.). The test strip vial contains a drying agent.

Performance Characteristics

The performance of blue OneTouch

®

 Ultra

®

 Test Strips has been evaluated both in laboratory 

and in clinical tests.

3

Measurement Range:

 The measurement range of the OneTouch

®

 Ultra

®

 Family of Meters is 

1.1 to 33.3 mmol/L (20– 600 mg/dL).

System Accuracy:

 Diabetes experts have suggested that glucose meters should agree within 

0.83 mmol/L (15 mg/dL) of a laboratory method when the glucose concentration is lower 
than 4.2 mmol/L (75 mg/dL), and within 20% of a laboratory method when the glucose 
concentration is 4.2 mmol/L (75 mg/dL) or higher. Samples from 100 diabetic patients at 1 
clinical center were tested using both the OneTouch

®

 Ultra

®

2 System and the YSI Model 2300 

Glucose Analyzer (laboratory test).

3

 

System Accuracy Results for Glucose Concentrations <4.2 mmol/L (75 mg/dL)

System Accuracy Results for Glucose Concentrations 

4.2 mmol/L (75 mg/dL)

System Accuracy Results across the entire Glucose Range

Therefore, 96.3% of the total results obtained with the OneTouch

®

 Ultra

®

2 System achieved 

the goal suggested by the diabetes experts.

    Regression  Statistics: 

Samples were tested in duplicate on each of 3 test strip lots. Results 

indicate that the OneTouch

®

 Ultra

®

2 System compares well with a laboratory method.

 

Precision:

Within Run Precision  

(100 venous blood tests per glucose level) 

Results show that the greatest variability observed between test strips when tested 
with blood is 2.5% or less.
Data generated using OneTouch

®

 Ultra

®

2 Meter. OneTouch

®

 Ultra

®

2 is representative 

of the OneTouch

®

 Ultra

®

 Family of Meters (OneTouch

®

 Ultra

®

, OneTouch

®

 Ultra

®

2, 

OneTouch

®

 UltraMini

®

, OneTouch

®

 UltraSmart

®

) and the OneTouch

®

 Ping

 System.

IMPORTANT:

 For complete operating instructions and other important technical 

information, refer to the User Guide that came with your system. IF YOU HAVE 

QUESTIONS ABOUT THE USE OF ANY LIFESCAN PRODUCT, PLEASE CONTACT the 

OneTouch

®

 Customer Care Line at 1 800 663-5521. If you cannot reach Customer 

Service, contact your healthcare professional for advice.

References

1.  Beaser, R.S. and Hill, Joan: The Joslin Guide to Diabetes. New York: Simon and Schuster 

(1995), p. 158.

2.  Canadian Diabetes Association Clinical Practice Guidelines Expert Committee. Canadian 

Diabetes Association 2008 Clinical Practice Guidelines for the Prevention and Management 
of Diabetes in Canada. CJ Diabetes 2008:32 (suppl 1) S11, S62.

3. Data on file.

 OUR COMMITMENT TO YOU: 

Our goal is to provide you with quality healthcare products and dedicated customer 

service. If you are not fully satisfied with this product, we offer a No-Risk, Money-

Back Guarantee and will refund your purchase within 30 days. Please contact us 

regarding a refund or with questions about the use of any OneTouch

®

 products. For 

assistance, please contact the OneTouch

®

 Customer Care Line at 1 800 663-5521.  

If you cannot reach Customer Service, contact your healthcare professsional  

for advice. 

Covered by one or more of the following U.S. patents: 5,708,247, 5,951,836, 6,241,862, 
6,284,125, and 7,112,265. Use of these test strips and associated monitoring device is 
protected under the following U.S. patents: 6,413,410, 6,733,655, 7,250,105. Purchase 
of the associated monitoring device does not act to grant a use license under these 
patents. Such a license is granted only when the associated monitoring device is used with 
OneTouch

®

 Ultra

®

 Test Strips. No test strip supplier other than LifeScan, Inc. is authorized to 

grant such a license. The accuracy of results generated with LifeScan meters using test strips 
manufactured by anyone other than LifeScan has not been evaluated by LifeScan.

Within 

± 0.28 mmol/L  

(5 mg/dl)        

Within 

± 0.56 mmol/L 

(10 mg/dl)          

Within 

± 0.83 mmol/L 

(15 mg/dl)     

Percent (and number) of meter results that 
match the laboratory test

48.8% (41/84)

84.5% (71/84)

100.0% (84/84)

Within  ±  5%        Within  ±  10%          Within  ±  15%      Within ± 20%

Percent (and number) of meter results that 
match the laboratory test

38.0% 

 (196/516)

68.0%  

(351/516)

88.2% 

 (455/516)

95.7% 

 (494/516)

Within ± 0.83 mmol/L  

(15 mg/dL) or ± 20%

Percent (and number) of meter results that 
match the laboratory test

96.3% (578/600)

Number  

of  

Subjects        

Number  
of  Tests          

Slope      Intercept

mmol/L 
(mg/dL)

95% CI Slope

95% CI Intercept

mmol/L (mg/dL)

Std. Error 

(Sy.x)

mmol/L 
(mg/dL)

R

2

100

 600

0.972

-0.012

(-2.657)

0.958 to 0.985

-0.168 to 0.145

(-5.415 to 0.102)

1.000 

(17.743)

0.973

Target 

Glucose
mmol/L
(mg/dL)

Mean 

Glucose
mmol/L
(mg/dL)

Standard 
deviation

mmol/L
(mg/dL)

Coefficient 

of 

 variation

(%)

2.2

(40)

2.28  

(41.0)

0.057 
(1.02)

2.50

5.6

(100)

5.40  

(97.4)

0.096 
(1.74)

1.78

7.2

(130)

6.70 

(120.7)

0.116 
(2.10)

1.74

11.1

(200)

11.15 

(200.9)

0.160 
(2.87)

1.43

16.7

(300)

16.96 

(305.6)

0.197 
(3.55)

1.16

Glucose

Levels

Mean 

Glucose
mmol/L
(mg/dL)

Standard 
deviation

mmol/L
(mg/dL)

Coefficient 

of 

 variation

(%)

LOW

2.59 

(46.6)

0.056 
(1.01)

2.18

MID

6.39 

(115.1)

0.121 
(2.19)

1.90

HIGH

19.47 

(350.8)

0.304 
(5.48)

1.56

Total Precision (200 control solution tests 

per glucose level) 