Accuracy
problem
cont’d
Scale gives
different
readings for the
same load.
Cable problem.
Bad connections cause changes when moving the
scale.
Bend and move the cable briskly, especially where
the cable is moving continuously while lifting.
While doing so, look at the display to see if it reacts
to the movements.
Potentiometer
problem.
Move the board and put pressure with fingers on
the potentiometers. If you see no change on the
screen, the potentiometer is good. If you see
changes on the screen, the potentiometer is bad
and needs to be replaced. Do not touch the contact.
CONDITION
SUB-
CONDITION
POSSIBLE
CAUSE
ACTION